Skip to content
R script that automates “iterative exclusion” (IE) experiments for Thermo instruments
R
Branch: master
Clone or download
Fetching latest commit…
Cannot retrieve the latest commit at this time.
Permalink
Type Name Latest commit message Commit time
Failed to load latest commit information.
IEomics.R
IEomics_Manual.docx
README.md

README.md

IE-Omics

IE-Omics is a R script that automates “iterative exclusion” (IE) experiments for Thermo instruments. This script can easily be modified for use with other instruments. IE-Omics applies Iterative exclusion to traditional data-dependent tandem mass spectrometry top N experiments (ddMS2-topN), where the N most intense ions detected in a full scan are selected for fragmentation. In IE, previously selected ions for fragmentation using ddMS2-topN are excluded from selection in sequential injections. Theoretically fragmentation of all ions above a certain intensity threshold can be acquired. Applying this technique in our laboratory to Red Cross plasma and substantia nigra, we were able to identify up to 69 % more lipids than using traditional data-dependent acquisition.

Jeremy P. Koelmel, Nicholas M. Kroeger, Emily L. Gill, Candice Z. Ulmer, John A. Bowden, Rainey E. Patterson, Richard A. Yost, Timothy J. Garrett: Expanding lipidome coverage using LC-MS/MS data-dependent acquisition with automated exclusion list generation. Journal of the American Society for Mass Spectrometry. (2017) 28: 908. doi:10.1007/s13361-017-1608-0