Changes to the table that describes how modules react when the e-stop is engaged. Adds new rows for the HEPA/UV module and plate reader. Removes single bullet item lists (might be an artifact from conversion to mkdocs). Replaces markdown w/ html for the table.
Manual says flex status light turns red. It does not. The light turns yellow, solid (not flashing or pulsing). Stays yellow until you reset the e-stop. Last minute change from checking.
Needed a last minute push re: status light info. Will change to regular PR. |
Codecov ReportAll modified and coverable lines are covered by tests ✅
Additional details and impacted files@@ Coverage Diff @@
## edge #18857 +/- ##
=======================================
Coverage 24.34% 24.34%
=======================================
Files 3312 3312
Lines 289304 289304
Branches 30805 30805
=======================================
Hits 70432 70432
Misses 218851 218851
Partials 21 21
Flags with carried forward coverage won't be shown. Click here to find out more. 🚀 New features to boost your workflow:
|
There was a problem hiding this comment.
Choose a reason for hiding this comment
The reason will be displayed to describe this comment to others. Learn more.
Change to HTML table is a warranted improvement here. A couple prose comments and questions.
Thanks for the suggestions. Applying now. Co-authored-by: Ed Cormany <edward.cormany@opentrons.com>
Thermocycler light turns red, not yellow.
Changes to the table that describes how modules react when the e-stop is engaged. Adds new rows for the HEPA/UV module and plate reader. Removes single bullet item lists (might be an artifact from conversion to mkdocs). Replaces markdown w/ html for the table.
Manual says flex status light turns red. It does not. The light turns yellow, solid (not flashing or pulsing). Stays yellow until you reset the e-stop. Last minute change from checking.
Thanks for the suggestions. Applying now. Co-authored-by: Ed Cormany <edward.cormany@opentrons.com>
Thermocycler light turns red, not yellow.
b0579e9
to
35c5ed4
Compare
…pentrons into estop-docs-revisions
Overview
Revisions to chapter 3, System Description, Emergency Stop Pendant section. Changes update the table that lists how different modules respond to E-stop activation:
JIRA RTC-580
Test Plan and Hands on Testing
Checked with SMEs and testing the robot and e-stop itself.
Changelog
Text changes to a table, contents and code.
Review requests
The usual process.
Risk assessment
Low.