You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
Recently our VD13D Prisma has sporadically crashed when running the Lifespan HCP rsfMRI protocol with R015 of the MB EPI sequence. This occurs in about 20% of the scans, in the middle of the run. No software reboot is required to continue scanning. The error dialog is pasted below.
This does not occur for other MB protocols (e.g. the HCP dwi). Reducing the number of slices by one MB factor (i.e. by 8) appears to resolve the issue.
Siemens is aware of the problem, but has not been able to fix it, despite swapping out the gradient amp.
Has anyone else encountered this?
The text was updated successfully, but these errors were encountered:
Recently our VD13D Prisma has sporadically crashed when running the Lifespan HCP rsfMRI protocol with R015 of the MB EPI sequence. This occurs in about 20% of the scans, in the middle of the run. No software reboot is required to continue scanning. The error dialog is pasted below.
This does not occur for other MB protocols (e.g. the HCP dwi). Reducing the number of slices by one MB factor (i.e. by 8) appears to resolve the issue.
Siemens is aware of the problem, but has not been able to fix it, despite swapping out the gradient amp.
Has anyone else encountered this?
The text was updated successfully, but these errors were encountered: