- Input is a 180-degree sinogram.
- An estimated 360-degree sinogram is generated by: flipping the 180-degree sinogram, shifting horizontally the sinogram to an estimated center-of-rotation (COR) position, and appending the resulting image to the original 180-degree sinogram.
- Metric of the estimated 360-degree sinogram, which is the mean of values in the double-wedge region in the Fourier space, is calculated.
-> The estimated COR corresponding to the minimum of the measured metrics is the best COR.
Figure 1. Explanation of how VoCentering works.- preview : [] -> To select which sinograms are used to calculate the COR. Users have to change this parameter, otherwise the plugin will calculate COR for every sinogram which is computationally expensive. Examples:
- preview : [:,mid,:] -> Calculate the COR of a tomographic dataset using the middle sinogram.
- preview : [:,50:end:100,:] -> Calculate the COR using sinograms having indices from 50 to the last with the step of 100.
- preview : [:,mid-10:mid+10,:] -> Calculate the COR using 20 sinograms around the middle sinogram.
- search_area : [-50, 50] -> Shifting range (in pixel) of the bottom sinogram (Fig. 1(b)) to search for the best COR. Example:
- search_area : [-150, 150] : Shifting the bottom sinogram from left (-150) to right (+150) with respect to the horizontal centre of the sinogram image.
- start_pixel : None -> Starting point for the search in parameter 2). If the value is None, the horizontal centre of the image is used. Example:
- start_pixel : 1100
- ratio : 0.5 -> To define the wedge angle of the mask (Fig. 1d). The value can be adjusted using the ratio between the size of a sample and the horizontal field-of-view of a detector. However, it's insensitive. The value of 0.5 works in most of cases. Examples:
- ratio : 0.2 : If a sample is very small with respect to the detector field-of-view.
- ratio : 2.0 : If a sample is bigger than the detector field-of-view.
- search_radius : 6 -> Searching range (in pixel) from the left (-6) to the right (6) of the coarse estimated COR to find the best COR with sub-pixel accuracy.
- step : 0.5 -> Sub-pixel shifting step of the bottom sinogram (Fig. 1(b)). This corresponds to the accuracy of the estimated COR of 0.25 pixel (= 0.5 / 2).
- broadcast_method : median -> Method for broadcasting the calculated COR to the whole dataset. For example, in the example above, a user choose to calculate the COR using 20 sinograms around the middle sinogram, then use the median value of the results to assign to all sinograms (e.g. 2000 sinograms). Example:
- broadcast_method : linear_fit -> Apply a linear fit to the calculated CORs, then broadcast to all sinograms.
- row_drop : 20 -> To ignore the values around the middle row of the Fourier of the sinogram (Fig. 1(d)).
- average_radius : 5 -> To improve the signal-to-noise ratio of low-contrast sinograms. The plugin averages neighbouring sinograms (+/- 5) around the current sinogram and uses the result for calculating the COR. This requires that the preview parameter (1) needs to be changed to provide these neighboring sinograms.
- Users often just need to adjust parameter 1) and 2). Default values of other parameters works in most of cases (~98%).
- About the order of the plugin in the data processing pipeline:
- Vocentering should be used before any plugin which blurs sinograms such as PaganinFilter, FresnelFilter, MedianFilter, or RavenFilter. Blurred sinograms affect the accuracy of the Vocentering plugin.
- The plugin must be used after the DistortionCorrection plugin, because the distortion alters the CORs from the top sinogram to the bottom sinogram in a nonlinear way.
- If a sample is too small compared to the width of the image, adjust the ratio (4) to something ~0.1 -> 0.3.
- If a sample is much bigger than the width of the image, adjust the ratio (4) to something ~ 2.0 -> 5.0.
- If projection images are slightly tilted, use 'linear_fit' for parameter (8).
- If images are at very low-contrast, increase parameter (10) and parameter (1) correspondingly.