You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
RobertGawron opened this issue
Aug 16, 2019
· 1 comment
Assignees
Labels
bugSomething isn't workinghardwareTopics that require changes in HW or needs HW interactions (oscilloscope, multimeter, etc.).softwareTopics that require changes in SW.
Probably ADC issue since no spikes are visible on oscilloscope
The text was updated successfully, but these errors were encountered:
RobertGawron
added
bug
Something isn't working
hardware
Topics that require changes in HW or needs HW interactions (oscilloscope, multimeter, etc.).
software
Topics that require changes in SW.
labels
Aug 16, 2019
Adding 68k resistor between U33B output and ground makes issues less visible but doesn't solve it.
Before it 1 in 2 measurements, after it's 1 in 10 measurements.
RobertGawron
changed the title
Resolve output voltage spikes when no sample is present
Resolve output voltage spikes when measuring background level
Aug 20, 2019
bugSomething isn't workinghardwareTopics that require changes in HW or needs HW interactions (oscilloscope, multimeter, etc.).softwareTopics that require changes in SW.
Probably ADC issue since no spikes are visible on oscilloscope
The text was updated successfully, but these errors were encountered: