Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

Volume calculations are off by factor of 2 #4

Open
cbonsig opened this issue Jun 6, 2017 · 0 comments
Open

Volume calculations are off by factor of 2 #4

cbonsig opened this issue Jun 6, 2017 · 0 comments

Comments

@cbonsig
Copy link
Contributor

cbonsig commented Jun 6, 2017

In the 17.06 release, a symmetry factor is 16 is used in the volumetric calculations: https://github.com/confluentmedical/nitinol-design-concepts/blob/master/125-volumetric-analysis/postprocessFEA.R#L57

This assumes that FEA output represents 1/16th of the full component. In fact, the symmetry factor should by 32. The model represents 1/32nd of the full component, and a symmetry boundary condition is applied such that the model represents 1/16th only when mirrored.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment