- CO2_CONC ... CO2 concentration density, molar density (mmol m-3)
- H2O_CONC ... H2O concentration density, molar density (mmol m-3)
- T_LI7500 ... Ambient temperature measured in the control box (deg C)
- P_LI7500 ... Atmospheric pressure measured in the control box (kPa)
- COOLER_V ... Cooler voltage
- GA_DIAG_VALUE ... IRGA diagnostic value
- same as GA_DIAG_CODE found in other datablocks
- The cell diagnostic value is a 1 byte unsigned integer (value between 0 and 255, or in binary
notation e.g.
00111001
) with the following bit map (in order of how the code reads it, orig bit position in brackets):- (7; first digit) CHOPPER: 1 = OK
- (6; second digit) DETECTOR: 1 = OK
- (5; third digit) PLL: 1 = OK
- (4; fourth digit) SYNC: 1 = OK
- (3,2,1,0; last four digits) AGC: 100% = bad (generally true, but check data files for exceptions),
automatic gain control, window dirtiness, signal strength. However, this value can never reach 100%
according to the binary notation, because AGC corresponds to the 4 right-most digits of the byte,
and if these are e.g.
1111
than that corresponds to decimal15
, which is multiplied by6.25
to yield AGC =93.75%
.
- 'output': 1 means that the var is written to the output stream of this data block, i.e. included in the output file.
- LI7500_STATUS ... Status of IRGA data aquisition, see Table 7 in WE's sonicread.pdf
- same as STATUS_CODE found in other datablocks
- octal value converted to integer yields:
- 0 .. Status OK, no problems (octal 0000)
- 20 .. IRGA did not respond (0020)
- 40 .. Status OK, old data used (0040)
- 200 .. not OK, IRGA data are missing (0200)
- This datablock does not need conversion because it is already logged in ASCII format.
- Datablock originates from the new logging script
rECord
, which will replace the old logging scriptsonicread
. - This datablock was originally used for
the
rECord
test phase at the research site CH-FRU in 2023. - More info about
sonicread
can be found indocs/sonicread_20190503.pdf
.
- Note in case
AGC
is needed: that since these files are not converted inbico
, theAGC
value from the IRGA has to be extracted fromGA_DIAG_VALUE
with other methods, e.g., by using a separate script.