Planar configuration for RGB images #31

Closed
jap1968 opened this issue May 2, 2013 · 0 comments
Closed

Planar configuration for RGB images #31

jap1968 opened this issue May 2, 2013 · 0 comments
Assignees
Labels
enhancement New feature or request
Milestone

Comments

@jap1968
Copy link
Contributor

jap1968 commented May 2, 2013

The problem with the RGB ultrasound image in your samples is related to the Dicom Element "Planar Configuration" (0028, 0006) having the value = 1

All the sample images used on my tests have this value set to "0", i.e. they were using the usual RGB ordering. Your sample image is the first one that I see having planar configuration = 1.

The particular bit ordering for planar configuration = 1 is the cause of the curious (and wrong) mosaic pattern displayed when trying to display the image.

You can find additional information here:
http://www.medicalconnections.co.uk/kb/Planar_configuration
https://www.dabsoft.ch/dicom/3/C.7.6.3.1.3/