{% hint style="warning" %} Advanced Users Only:
Currently called by the GUI
Use GUI for calibrations in most cases {% endhint %}
The GUI will run calibrations automatically upon completion of the calibration protocols. However, you can still manually run a calibration if you would like to change calibration settings. Can be useful if your GUI calibration fails for some reason.
Located in /dpu/calibration/.
- You have run calibration and logged raw values already
- You are in your DPU virtual environment in command line. This is set up when you install the DPU.
Mac
python3 calibration/calibrate.py -a <ip_address> -g
For Windows, use py instead of python for all commands.
python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t linear -f <name_after_fit> -p temp
OD135
python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t sigmoid -f <name_after_fit> -p od_135
OD90 (Check to ensure mode is configured properly)
python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t sigmoid -f <name_after_fit> -p od_90
3D FIT (Check to ensure mode is configured properly)
python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t 3d -f <name_after_fit> -p od_90,od_135