Maximum bolus is ignored #33
Labels
Comments
I tested this by setting my max to 0.1u in app and giving myself a 0.2u bolus and it gave the bolus as well. |
Yep – |
Related to this would be great to validate carb entry, yesterday Sam by mistake keyed 5523 gr vs 55 and Loop suggested 553 units. |
loudnate
added a commit
that referenced
this issue
Sep 28, 2016
erikdi
pushed a commit
to erikdi/Loop
that referenced
this issue
Feb 10, 2019
Xcode 7.3 / Swift 2.2 support
ps2
pushed a commit
that referenced
this issue
Jan 18, 2020
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
I have maximum bolus in Loop set to 1U (and 6U on the pump itself) and was just able to do a 1.5U bolus from Loop.
The text was updated successfully, but these errors were encountered: