Notifications
You must be signed in to change notification settings - Fork 122
/
VESUVIO_Parameters.xml
131 lines (102 loc) · 3.34 KB
/
VESUVIO_Parameters.xml
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
<?xml version="1.0" encoding="UTF-8" ?>
<parameter-file instrument = "VESUVIO">
<component-link name = "VESUVIO">
<!-- The monitor spectrum -->
<parameter name="monitor_spectrum">
<value val="1"/>
</parameter>
<!-- Back scattering spectrum numbers -->
<parameter name="backward_scatter_spectra" type="string">
<value val="3-134"/>
</parameter>
<!-- Forward scattering spectrum numbers -->
<parameter name="forward_scatter_spectra" type="string">
<value val="135-198"/>
</parameter>
<!-- Maximum TOF for loaded raw data (monitors use mon_tof_max below) -->
<parameter name="tof_max">
<value val="600"/>
</parameter>
<!-- Maximum TOF for loaded monitor data -->
<parameter name="monitor_tof_max">
<value val="700"/>
</parameter>
<!-- Constant that sets the monitor scale -->
<parameter name="monitor_scale">
<value val="1000"/>
</parameter>
<!-- Double-difference mixing parameter (beta) -->
<parameter name="double_diff_mixing">
<value val="0.28"/>
</parameter>
<!-- ======== Normalisation ranges. ========
These are different depending for the forward/backward scattering banks.
-->
<!-- Backward Scattering -->
<!-- Sum over lower range -->
<parameter name="backward_period_sum1" type="string">
<value val="300-400"/>
</parameter>
<!-- Sum over upper range -->
<parameter name="backward_period_sum2" type="string">
<value val="550-600"/>
</parameter>
<!-- Monitor normalisation -->
<parameter name="backward_monitor_norm" type="string">
<value val="600-700"/>
</parameter>
<!-- Normalisation by foil out -->
<parameter name="backward_foil_out_norm" type="string">
<value val="410-430"/>
</parameter>
<!-- ====================================================== -->
<!-- Forward Scattering -->
<!-- Sum over lower range -->
<parameter name="forward_period_sum1" type="string">
<value val="100-400"/>
</parameter>
<!-- Sum over upper range -->
<parameter name="forward_period_sum2" type="string">
<value val="550-600"/>
</parameter>
<!-- Monitor normalisation -->
<parameter name="forward_monitor_norm" type="string">
<value val="600-700"/>
</parameter>
<!-- Normalisation by foil out -->
<parameter name="forward_foil_out_norm" type="string">
<value val="400-450"/>
</parameter>
<!-- ======== Fitting parameters ========
These are currently at the instrument level but some may need
move to the component level
-->
<!-- Standard deviation of source-sample distribution (metres) -->
<parameter name="sigma_l1">
<value val="0.021"/>
</parameter>
<!-- Standard deviation of sample-detector distribution (metres) -->
<parameter name="sigma_l2">
<value val="0.023"/>
</parameter>
<!-- Standard deviation of detector theta distribution (radians) -->
<parameter name="sigma_theta">
<value val="0.028"/>
</parameter>
<!-- Final energy of neutrons (meV) (Currently common but should it be?) -->
<parameter name="efixed">
<value val="4908"/>
</parameter>
<!-- Lorentzian HWHM of the foil analyser energy (efixed) (meV) -->
<parameter name="hwhm_energy_lorentz">
<value val="24"/>
</parameter>
<!-- Gaussian standard deviation of the foil analyser energy (efixed) (meV)
Note that this is not the HWHM but the standard deviation unlike the
Lorentzian parameter -->
<parameter name="sigma_energy_gauss">
<value val="73"/>
</parameter>
<!-- ====================================================== -->
</component-link>