Click on the hamburger menu in the top left corner.
:alt: Hamburger menu
-
Click on Config Builder.
:alt: Config builder
-
After selecting the Diaconn G8 Pump click on the Settings icon (cog wheel).
:alt: Settings
-
Choose Selected pump.
:alt: Select pump
-
Select your insulin pump’s model number once it appears in the list.
:alt: Pump pairing
-
There are two options to check your model number:
- The last 5 digits of the SN number on the back of the pump.
- Click on O button > Information > BLE > Last 5 digits.
:alt: check model no.
-
Once you select your pump, a window appears asking for a pin code. Enter the pin number displayed on your pump to complete the connection.
:alt: PIN code
-
Once your pump is connected, click on the Bluetooth symbol to check the status and to synchronize logs.
:alt: Bluetooth status
What to do in the case of an unstable Bluetooth connection with the pump.
-
Click on the 3 dots button on the top right.
:alt: Preferences menu
-
Click on Exit.
:alt: Exit
-
Press and hold the Bluetooth button at the top for about 3 seconds.
:alt: Bluetooth button
-
Click on the Setting button on the paired Diaconn G8 Insulin pump.
:alt: Settings button
-
Unpair.
:alt: Unpair
-
Repeat the Bluetooth pairing process for the pump (see above).
- Config manager > pump > Diaconn G8 > Settings
- DIACONN G8 at the top> 3 dots button on the top right > Diaconn G8 Preferences
:alt: Diaconn G8 pump options
- If the Log reservoir change option is activated, the relevant details are automatically uploaded to the careportal when an “Insulin Change” event occurs.
- If the Log needle change option is activated, the relevant details are automatically uploaded to the careportal when a “Site Change” event occurs.
- If the Log tube change option is activated, the relevant details are automatically uploaded to the careportal when a “Tube Change” event occurs.
- If the Log battery change option is activated, the relevant details are automatically uploaded to the careportal when a “Battery Change” event occurs, and the PUMP BATTERY CHANGE button in the ACTION tab is deactivated. (Note: To change the battery, please stop all in-progress injection functions before proceeding.)
:alt: Diaconn G8 actions menu
- If you use extended bolus it will disable closed loop.
- See this page for details why extended bolus does not work in a closed loop environment.