Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

UHFLI default setting is not option selective #4

Open
chunyanusername opened this issue Oct 6, 2020 · 0 comments
Open

UHFLI default setting is not option selective #4

chunyanusername opened this issue Oct 6, 2020 · 0 comments

Comments

@chunyanusername
Copy link

chunyanusername commented Oct 6, 2020

In the Labber instrument server, after adding UHFLI, a path error message appears when starting the instrument. The screenshots are shown below.

The problem is that when starting the instrument, it loads the default setting in .ini file which includes the settings not only for a bared device but also for some options. In this case, the default setting includes setting with the MF option, i.e., set amplitude and frequency for 8 oscillators, but if the user has no MF option, a similar path as 'sigouts/0/amplitudes/0' will lead an error.

I have commented on the settings related to the MF option in .ini folder, have tested it and it works. However, this is just a temporary solution. I think the best way is that first identify what options are in the device and then load the default setting accordingly. I believe this issue also exists in other instruments.

UHFLI_Labber_output_interface
UHFLI_Labber_path_error_message

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment