Notifications
You must be signed in to change notification settings - Fork 0
/
CP-PCD-053-00.txt
109 lines (77 loc) · 4.89 KB
/
CP-PCD-053-00.txt
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
IHE Change Proposal
Tracking information:
|IHE Domain |Patient Care Devices |
|Change Proposal ID: |CP-PCD-053 |
|Change Proposal Status: |Ballot |
|Date of last update: |04/26/2011 |
|Person assigned: |John Rhoads |
Change Proposal Summary information:
|PIV Open Issues section |
|Submitter’s Name(s) and e-mail |Jeff Rinda |
|address(es): |(jeff.rinda@hospira.com) |
|Submission Date: |04/26/2011 |
|Integration Profile(s) affected: |Point-of-Care Infusion |
| |Verification (PIV) |
|Actor(s) affected: |None |
|IHE Technical Framework or |Technical Framework version 1.2 |
|Supplement modified: |(9/30/2010) |
|Volume(s) and Section(s) |Vol 1 – PIV section |
|affected: |Vol 2 – Section 3.3.4.1; also |
| |update reference to PIV |
| |application error table if |
| |possible |
| |PIV Application Error table |
| |(standalone document) |
|Rationale for Change: |
|Added a new application error message for care area or profile |
|mismatch; also add a more specific reference in the TF vol 2 (B.3 |
|ERR segment; ERR-5 section) about where the application error table |
|can be found on the IHE site (since this table is not included in |
|the TF). |
|Added a Limitations section to Vol 1 including an explanation for |
|how the PIV profile can support rate changes or titrations. |
|Added a statement to Vol. 2 about the use of information in the |
|PCD-03 transaction. |
Add the following row to the PIV application error table (Table 0533
Application Error Code. Insert a reference in Vol 2 about the existence
and location of this table (since it is not part of the TF).
|9021 |Care area or profile mismatch |Care area or profile |
| | |not cleared on pump; |
| | |or pump is set to a |
| | |different care area |
The following is a revision of text that was formerly contained in the Open
Issues section of the PIV supplement. This text should be incorporated
into TF Vol 1 as a “Limitations” section.
Limitations
1. The principal use case involves starting a new infusion or container.
Titrations or rate changes on a previously programmed infusion are
supported if one of the following applies:
a) The Infusion Order Programmer is able to provide the amount of the
volume currently present in the bag in the PCD-03 transaction (rather
than the ordered volume)
b) The BCMA application provides an acceptable workflow for the clinician
to adjust the volume to be infused directly on the pump from the
amount ordered to the amount currently present in the bag
c) The Infusion Order Consumer is able to interpret the PCD-03
transaction as a rate change without explicit identification as such
(since there is no element in the transaction that refers to whether
it is for a new bag vs. a rate change).
2. Patient-controlled analgesia (PCA) infusion devices are not supported.
3. The following therapies, modes, and methods of operation are not
supported:
a) Bolus, loading dose, multi-step, periodically repeating or ramp/taper
therapies
b) Delayed start mode
c) Programming concurrent or sequenced infusions on more than one pump
channel using a single transaction
4. PIV can be used to support administration of medications from a syringe
and/or using a syringe pump device if the Infusion Order Consumer does not
require the syringe size to be specified.
Add the following change toTF Vol 2 in Section 3.3.4.1
PCD-03 Communicate Infusion Order (RGV^O15^RGV_O15) static definition
The PCD-03 Communicate Infusion Order message is used to communicate
infusion data from an Infusion Order Programmer (IOP) to an Infusion Order
Consumer (IOC).
Since the IOC is typically a gateway rather than an infusion pump, all of
the information specified in the PCD-03 Communicate Infusion Order
transaction is not necessarily provided to or used to program the device.