IOC: Linkam95 does not send 'T' command first #1509

Closed
AdrianPotter opened this issue Aug 23, 2016 · 4 comments
Closed

IOC: Linkam95 does not send 'T' command first #1509

AdrianPotter opened this issue Aug 23, 2016 · 4 comments

Comments

@AdrianPotter
Copy link

AdrianPotter commented Aug 23, 2016

The Linkam T95 manual states that the first command sent to the device must be the 'T' command. At the moment, the way scanning is set up, the command to set the pump mode goes first, and is thus ignored. Please make it so 'T' is the first thing that is sent.

@KathrynBaker
Copy link
Member

This might need the same solutions as the Danfysik and Genesys PSUs – they actually send a command in the st.cmd before the db files are loaded for very similar reasons.

@AdrianPotter
Copy link
Author

@AdrianPotter
Copy link
Author

@AdrianPotter
Copy link
Author

Thanks @KathrynBaker for the input. That would definitely have been a solution. I note that the McLennan also does a similar thing. In the end @IsabellaRey helped me do it via some PV trickery.