Autosampling titrator for measuring of dKH level in saltwater aquarium
Join the Arduino Aquarium titrator Group Chat if you have any questions/feedback or would just like to be kept up to date with the project progress.
-
Register on https://labaqua.net/ get token on email
-
To get Telegram notification you should generate new Bot and get an Access Token. Talk to BotFather and follow a few simple steps described here.
-
Connect device to PC throughout USB, open COM-port terminal (monitor_speed: 115200) and do next steps
-
Set the PASSWORD for your WiFi, e.g. wifipasw_YourWiFi#YourPassword
-
Set the token from email, ukey_1Egu-78jt7NpQ7gau5d778a299e0a43,67773777
-
Set the telegram bot access token, e.g. bottoken_123789548:RRRRRGseAZ1ikRIhpjGa1-asdfghGVWqwers
-
Get your telegram ID and set up it to the device so only you have access, settlgrmid_123456789 Restart the device, if all is done fine you see the notification "KH station started" on your chatbot.
- Open chatbot, click "Calibrations"
- Select "Calibrate pumps"
- Setup the desired hose-tip over the weighed vessel and click "Run pump test"
- After ending pumping (beeping three times) measure the mass of the pumped liquid. If the mass is out of the admissible¹ interval the real pumped mass should be sent to the device for correction. Click "Set real mass", after that enter the mass, e.g. 24.87 Repeat calibrations until getting admissible¹ results.
¹ reagent 10.00±0.05 g, water 25.00±0.05 g