Over pressurization #39

Closed
fireb0lt opened this issue Mar 20, 2020 · 3 comments
Closed

Over pressurization #39

fireb0lt opened this issue Mar 20, 2020 · 3 comments

Comments

@fireb0lt
Copy link

I don't think that 20cm h2o is very high, nor would cause any lung trauma. My personal NIV mask setup tops out at 25cm and isn't even fully inflating my lungs. Cough assist machines go into the 60s-70s of pressure, even if it is for brief periods.

All this to say, we likely will have problems getting high enough pressures, don't need to really worry about over pressurization. As a FYI, I'm from Washington State where Covid-19 is out of control. My pulmonologist is one of the top in the nation (Dr. Joshua Benditt) and we talked about too high of pressures on my ventilator. He wasn't concerned.

@jcl5m1
Copy link
Owner

jcl5m1 commented Mar 20, 2020

interesting, i get mixed info on this. I agree 20cmh2o doesn't feel high, but 45cmh2o kind of hurts as a healthy individual. I don't know how sensitive COVID patients can get, so I have to defer to the experts. I'd be interested in Joshua's read on the following:

https://github.com/jcl5m1/ventilator/wiki/Notes-from-chatting-with-a-pulmonologist

@nasocializes nasocializes assigned fireb0lt and unassigned fireb0lt Mar 25, 2020
@nasocializes
Copy link
Collaborator

Johnny: it seems like going above 20cm requires additional sensors to ensure safety. we'll continue to use 20cm as a cut off guidance until we have more capabilities.

Thank you! Closing for now to clean up the list.

@bradtem
Copy link

bradtem commented Mar 25, 2020

Consulting with some doctors who do ventilation, pressures are high in ARDS. The alveoli are closed up and flooding. Lungs are not compliant. As such, ARDS ventilation is done with a very high base pressure (known as PEEP.) Doctors are setting this to 16cm in may cases. This is the lowest pressure, for exhalation. The peak pressure varies, enough to deliver the required volume of air (or rather oxygen.) Because the alveoli are so stressed, it is desired to deliver a very low volume of air, which means having a very high concentration of oxygen. I have heard reports patients start at 95%, which can't be sustained because it is toxic but they get them going this way. If the patient starts at 16cm, I suspect a peak pressure of 20cm will not help many patients. Of course there are patients not needing the 16cm, or the higher peaks. Also, you end up pumping nearly pure oxygen, which clearly is a risky thing to do near heat sources. Just about anything can burn in that.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment