Skip to content
/ labdaq Public

tools to control daq hardware for ephys or stimulation

License

Notifications You must be signed in to change notification settings

jcouto/labdaq

Folders and files

NameName
Last commit message
Last commit date

Latest commit

 

History

49 Commits
 
 
 
 
 
 
 
 
 
 
 
 

Repository files navigation

 _       _         _             
| |     | |       | |            
| | __ _| |__   __| | __ _  __ _ 
| |/ _` | '_ \ / _` |/ _` |/ _` |
| | (_| | |_) | (_| | (_| | (_| |
|_|\__,_|_.__/ \__,_|\__,_|\__, |
                              | |
                              |_|
 https://github.com/jcouto/labdaq

Tools to control daq hardware for ephys or stimulation. This aims to make easy tasks IO tasks simple and to encompass a broad range of applications.

This is currently under development.

Supported features:

  • interfacing with an AXON 700B amplifier using telegraph input
  • stimuli generation
  • GUI for patch clamping (seal test, synchrounous acquisition)
  • support for NI devices (tested with the PCI 6229)

Usage:

The first time it is ran it creates a configuration file in $HOME/labdaq/default/prefs.json

This configuration file specifies 3 things:

  • channel_modes the available modes for all channels.
  • channels list of channels.
  • recorder options which specifies the format and the datapath

channel_modes

These can have different units and conversion factors. Each mode can have different settings for input and output.

Example:

    "channel_modes": {
        "analog": {
            "description": "regular_output",
            "input_conversion": 1,
            "input_units": "V",
            "output_conversion": 1,
            "output_units": "V"
        },
        "cc": {
            "description": "current clamp",
            "input_conversion": 100,
            "input_units": "pA",
            "output_conversion": 5e-05,
            "output_units": "mV"
        },
        "ttl": {
            "description": "TTL digital logic",
            "input_conversion": 1,
            "input_units": "na",
            "output_conversion": 1,
            "output_units": "na"
        },
        "vc": {
            "description": "voltage clamp",
            "input_conversion": 1000,
            "input_units": "pA",
            "output_conversion": 0.05,
            "output_units": "mV"
        }
    }

channels

Each channel must specify:

  • the name
  • the device to use (use the driver:device notation)
  • the type of channel (e.g. analog_output)
  • the sampling rate acq_rate (in samples per sec)
  • a list of valid channel modes and
  • the acquisition range (in Volt)

Example:

This is for patch clamping with the axon 700B.

    "channels": [
        {
            "acq_rate": 40000,
            "channel": "ai0",
            "device": "ni:Dev2",
            "modes": [
                "vc",
                "cc"
            ],
            "name": "ephys_in",
            "range": [
                -10,
                10
            ],
            "type": "analog_input"
        },
        {
            "acq_rate": 40000,
            "channel": "ao0",
            "device": "ni:Dev2",
            "modes": [
                "vc",
                "cc"
            ],
            "name": "ephys_out",
            "range": [
                -10,
                10
            ],
            "type": "analog_output"
        },
        {
            "acq_rate": 40000,
            "channel": "port0",
            "device": "ni:Dev2",
            "modes": [
                "ttl"
            ],
            "name": "opto_out",
            "type": "digital_port"
        },
        {
            "acq_rate": 40000,
            "channel": "ao2",
            "device": "ni:Dev2",
            "modes": [
                "analog"
            ],
            "name": "analogstim_out",
            "range": [
                -5,
                5
            ],
            "type": "analog_output"
        },
        {
            "channel": "ai1",
            "device": "ni:Dev2",
            "modes": [
                "analog"
            ],
            "name": "ephys_mode",
            "type": "axon200B_mode"
        }
    ],

Experiment protocol files

This is meant to allow reproducible experiments without loosing flexibility. It is still being developed.

Protocol files describe what is the output of analog output files, the intertrial interval and how other parameters.

These are simple text files. Available options:

  • ntrials=X number of trials
  • iti=X inter-trial interval in seconds
  • analog_stim=FILE1,FILE2,FILE3 assigns waveform files to each analog output channel
  • digital_stim=FILE1,FILE2,FILE3 assigns waveform files to digital output channels

Any other variable can be defined and it will be replaced in the waveform files if it is encompassed by curly brackets (like in the python strings format method).

interfacing with labcams

To trigger with labcams (https://bitbucket.org/jpcouto/labcams/src/master/) add the labcams=ADDRESS:PORT line to the expprot file. Labcams will be triggered and the files will be named accordingly.

Using the command line

Launch with labdaq. More options will be described here.

Installation

Clone the repository. Install from the code folder using python setup.py develop

Dependencies:

  • numpy
  • scipy
  • matplotlib
  • pyqt5
  • pyqtgraph
  • pandas
  • hdf5
  • nidaqmx (for using NI cards)

More on the installation will come here.