Merge pull request #175 in radformation/pylinac from docs/cbct_acquis…
Browse files Browse the repository at this point in the history
…ition_warning to master

add cbct acquisition warning

Approved by Randy Taylor
  • Loading branch information
jrkerns committed Feb 23, 2023
2 parents b059146 + b2f0344 commit b83e51c
Showing 1 changed file with 14 additions and 0 deletions.
14 changes: 14 additions & 0 deletions docs/source/cbct.rst
Original file line number Diff line number Diff line change
Expand Up @@ -69,6 +69,20 @@ Or generate a PDF report:
cbct.publish_pdf('mycbct.pdf')
Image Acquisition
-----------------

Acquiring a scan of a CatPhan has a few simple requirements:

#. The field of view must be larger than the phantom diameter + a few cm for clearance.
#. All modules must be visible.

.. warning::

This can cause strange results if not all modules are scanned. Furthermore, aligning axially to the white dots
on the side of the catphan will not catch the inferior modules on a typical CBCT scan. We suggest
aligning to the center of the HU module (the module inferior to the white dots) when acquiring via CBCT.