Calibrations

{% hint style="warning" %} Before beginning calibrations all Software Installation and Network Configuration should be complete. {% endhint %}

About

In order to appropriately interpret data from and actuate physical elements connected to the eVOLVER framework, a relationship must be established between the electrical signals generated by attached sensors and the physical phenomenon they are measuring. This is done by manually measuring values on a controlled set of data via some gold-standard assay or measurement device and comparing these data to the eVOLVER generated data. By doing this across a range of different values, a function can be fit to this relationship to allow eVOLVER to interpret data accurately.

For more information see here.

Procedure

Calibrations should proceed in the following order:

  1. Temperature Calibration
  2. Optical Density Calibration
  3. Pump Calibration

{% hint style="warning" %} If the GUI fails to calibrate you can try to use calibrate.py manually. {% endhint %}