Skip to content

MakAir Control UI v2.2.0

Compare
Choose a tag to compare
@valeriansaliou valeriansaliou released this 02 Feb 16:21
· 52 commits to master since this release
630bd8a
  • Added a patient configuration modal, that is shown on startup, asking for the patient size in order to apply recommended ventilation parameters automatically [@valeriansaliou].
  • Added a peak pressure alarm [@valeriansaliou].
  • Added support for more firmware-related exceptions, eg. sensor error, watchdog fired, etc. [@valeriansaliou].
  • Improved the battery SoC estimation, which is still done without the help of a BMS [@valeriansaliou].
  • The animated breath indicator now uses the peak pressure alarm setting as a maximum reference, if configured [@valeriansaliou].
  • The user interface and the appearance of some buttons have been improved, based on user tests [@valeriansaliou].