Skip to content

Commit

Permalink
Add function for WISH re #6764
Browse files Browse the repository at this point in the history
This function needs a second argument to specify the panel to calibration, because each calibration run deals with only one panel.
Also the arguments are string rather than numbers.

Signed-off-by: Karl Palmen <karl.palmen@stfc.ac.uk>
  • Loading branch information
KarlPalmen committed Mar 26, 2013
1 parent 474f59a commit b01b9bc
Showing 1 changed file with 64 additions and 0 deletions.
Original file line number Diff line number Diff line change
@@ -0,0 +1,64 @@
#
# TUBE CALIBRATION DEMONSTRATION PROGRAM FOR WISH
#
# Here we run the calibration of WISH panel03 using a simple CalibrateWish function.
#
from mantid.api import WorkspaceFactory # For table worskspace of calibrations
from ideal_tube import * # for ideal tube
from tube_calib_fit_params import * # To handle fit parameters
from tube_calib import * # For tube calibration functions
from tube_spec import * # For tube specification class

def CalibrateWish( RunNumber, PanelNumber ):
'''
RunNumber is the run number as a string and including any leading zeros that would occur in the raw file name.
PanelNumber is the two-digit number of the panel being calibrated
'''
# == Set parameters for calibration ==
filename = 'WISH'+RunNumber+".raw"
CalibratedComponent = 'WISH/panel'+PanelNumber


# Get calibration raw file and integrate it
print "Loading",filename
rawCalibInstWS = Load(filename) #'raw' in 'rawCalibInstWS' means unintegrated.
CalibInstWS = Integration( rawCalibInstWS, RangeLower=1, RangeUpper=20000 )
DeleteWorkspace(rawCalibInstWS)
print "Created workspace (CalibInstWS) with integrated data from run and instrument to calibrate"

#Create Calibration Table
calibrationTable = CreateEmptyTableWorkspace(OutputWorkspace="CalibTable")
calibrationTable.addColumn(type="int",name="Detector ID") # "Detector ID" column required by ApplyCalibration
calibrationTable.addColumn(type="V3D",name="Detector Position") # "Detector Position" column required by ApplyCalibration

# Specify panel of WISH instrument
thisTubeSet = TubeSpec(CalibInstWS)
print "Calibrating",CalibratedComponent
thisTubeSet.setTubeSpecByString(CalibratedComponent)

# Give y-positions of slit points (gotten for converting first tube's slit point to Y)
iTube = IdealTube()
iTube.setArray([-0.41,-0.31,-0.21,-0.11,-0.02, 0.09, 0.18, 0.28, 0.39 ])

# Set fitting parameters
eP = [57.5, 107.0, 156.5, 206.0, 255.5, 305.0, 354.5, 404.0, 453.5]
fitPar = TubeCalibFitParams( eP, 2000, 32 )

print "Created objects needed for calibration."

# Get the calibration and put it into the calibration table
getCalibration( CalibInstWS, thisTubeSet, calibrationTable, fitPar, iTube)
print "Got calibration (new positions of detectors)"

#Apply the calibration
ApplyCalibration( Workspace=CalibInstWS, PositionTable=calibrationTable)
print "Applied calibration"

# == Save workspace ==
nexusName = "TubeCalibDemoWish"+PanelNumber+"Result.nxs"
SaveNexusProcessed( CalibInstWS, 'TubeCalibDemoWishResult.nxs',"Result of Running TubeCalibWishMerlin_Simple.py")
print "saved calibrated workspace (CalibInstWS) into Nexus file",nexusName

# ==== End of CalibrateWish() ====

CalibrateWish( '00017701', '03' )

0 comments on commit b01b9bc