Keithley 2600: support returning both voltage and current compliance limit errors #3339

Merged
merged 5 commits into from
Sep 9, 2021

Conversation

astafan8
Copy link
Contributor

@astafan8 astafan8 commented Sep 8, 2021

otherwise when in the SMU is used in current sourcing mode and the voltage hits the voltage compliance limit, the measurement_status property of the current/voltage parameters did not reflect that.

@astafan8 astafan8 added this to the 0.29.0 milestone Sep 8, 2021
@codecov
Copy link

codecov bot commented Sep 9, 2021

Codecov Report

Merging #3339 (60ebf61) into master (3c96817) will increase coverage by 0.00%.
The diff coverage is 100.00%.

@@           Coverage Diff           @@
##           master    #3339   +/-   ##
=======================================
  Coverage   66.20%   66.21%           
=======================================
  Files         220      220           
  Lines       29250    29253    +3     
=======================================
+ Hits        19365    19369    +4     
+ Misses       9885     9884    -1     

@astafan8 astafan8 enabled auto-merge (squash) September 9, 2021 13:59
@astafan8
Copy link
Contributor Author

astafan8 commented Sep 9, 2021

tested on the instrument and confirmed that it works.

@astafan8 astafan8 merged commit 7fa675f into master Sep 9, 2021
@astafan8 astafan8 deleted the astafan8-2600-compliance branch September 9, 2021 14:48