simonzihlmann
Copy link
Contributor

make frequency option detection compatible with multiple options installed on device. The old version of the driver resulted in a KeyError if multiple options were installed on the device.

I was unable to figure out from the manual / spec sheet if the frequency option will always be the first one in the string returned by the instrument. For the moment, all other options are ignored.

@codecov
Copy link

codecov bot commented Jul 1, 2022

Codecov Report

Merging #4342 (d3fc2fe) into master (c4983cd) will decrease coverage by 0.01%.
The diff coverage is 0.00%.

@@            Coverage Diff             @@
##           master    #4342      +/-   ##
==========================================
- Coverage   68.37%   68.36%   -0.02%     
==========================================
  Files         267      267              
  Lines       30961    30968       +7     
==========================================
  Hits        21170    21170              
- Misses       9791     9798       +7     

@jenshnielsen
Copy link
Collaborator

@simonzihlmann Thanks, could you add a news fragment as described in the pr template?

@jenshnielsen jenshnielsen mentioned this pull request Jul 4, 2022
@jenshnielsen
Copy link
Collaborator

bors merge

@bors
Copy link
Contributor

bors bot commented Jul 4, 2022

🕐 Waiting for PR status (Github check) to be set, probably by CI. Bors will automatically try to run when all required PR statuses are set.

@bors bors bot merged commit c42d7b9 into microsoft:master Jul 4, 2022
@@ -13,6 +13,27 @@ class N51x1(VisaInstrument):
def __init__(self, name: str, address: str, min_power: int = -144, max_power: int = 19, **kwargs: Any):
super().__init__(name, address, terminator='\n', **kwargs)

self._options = self.ask("*OPT?")
Copy link
Contributor

Choose a reason for hiding this comment

The reason will be displayed to describe this comment to others. Learn more.

the comment in the old code makes me think that the instrument response is a comma-separated list of numbers, right? if so, then it would've been better to split the returned string into an actualy list of strings with numbers. the if f_option in self._options check would still work, but would be more robust, so that e.g. option 50101 does not get mistakenly treated as option 501 (but of course likely option 50101 does not exist, so this all is likely not a problem in real life)

self._options = self.ask("*OPT?").split(",")