FastScanning.jl

CI codecov

Package for calibrating fast scanning with the fast OCPI microscope published in:

alt text

Please see the publication for the full rationale of this package. Below is a brief guide to use.

using FastScanning, ImagineInterface, Unitful
import Unitful: Hz, μm

rig = "ocpi-2"
pos_name = "axial piezo"
#set stacks/s (this will be twice the scan frequency of the piezo due to bidirectional imaging)
stack_rate = 10.0Hz
#lower limit of the piezo scan.  It's recommended not to start at the min position (0μm)
pstart = 50.0μm
#upper limit of the piezo scan.  It's recommended not to use the max position.
pstop = 260.0μm
#set z axis sample spacing
z_spacing = 5.0μm
#set amount of space to leave at edges of scan cycle
#i.e. if z_pad is 5.0μm and pstart:pstop is set as above,
#...then only the region from 55μm - 255μm will be imaged
z_pad = 5.0μm

coms = pos_commands(rig, pos_name, pstart, pstop, stack_rate;
                      sample_rate = 100000Hz, lp_cutoff=3.5*stack_rate)

The pos_commands function generates a set of commands that can be run by Imagine.jl for the first calibration step. No images are acquired in this step; we just record the response of the piezo to the high-speed command. Note that sample_rate and lp_cutoff are set to their default values above. lp_cutoff is the cutoff of the lowpass filter applied to the triangle wave command.

NOTE: Increasing stack_rate, lp_cutoff, or the piezo range (pstop-pstart) may result in commands that are not considered safe to run with Imagine because they may damage the piezo. In this case you will get an error when saving the commands to file or when trying to run them in the Imagine GUI.

The following code assumes the "coms" commands were run with the function run_imagine from the Imagine.jl package, and then the results of the experiment were loaded into the "recs" variable using ImagineInterface.jl.

pr = slicetiming_experiment(coms, recs, "488nm laser", "camera1", z_spacing, z_pad; subpixel=true)
st_coms = outputs(pr)

pr is an ImagineProcedure (see ImagineProcedures.jl) that generates calibration commands for the microscope and then later analyzes the results of running the commands (see below).

The following code assumes that the "st_coms" commands were saved to a file with ImagineInterface.jl and then run with the Imagine GUI (imaging a fluorescent bead sample), and the resulting image was loaded into the "img" variable.

fwd_lags, back_lags, slice_zs = process(pr, img)

fwd_lags and back_lags are vectors of temporal offsets (in seconds), with one offset for each slice in the "forward" and "reverse" stacks. These calibration values can be applied by subtracting them from the corresponding naive sensor-based exposure timings to determine the calibrated exposure timing. It's recommended to run this calibration just before performing an experiment. If any scan parameter changes then the calibration should be run again. slice_zs contains the precise depth at which each slice is acquired, respecting the pstart, pstop, z_spacing, and z_pad arguments provided by the user.

About

Determine camera/laser timing during fast OCPI acquisitions based on measured hardware responses.