Imagine

A graphical interface for recording with OCPI microscopes.

Companion paper

A paper explaining most of the implementation details and related hardware can be found here.

@article{Greer2019,
  title={Fast objective coupled planar illumination microscopy},
  author = {Greer, Cody and Holy, Timothy E.},
  journal={Nature Communications},
  volume={10},
  pages={4483},
  year={2019},
  URL = {https://doi.org/10.1038/s41467-019-12340-0},
  eprint = {https://www.nature.com/articles/s41467-019-12340-0.pdf},
  doi = {10.1038/s41467-019-12340-0}
}

Installation

We need to install below items

  • Imagine
  • Software development tools and libraries : Visual studio, QT, QWT, Boost, FFTW
  • Hardware related SDK and driver : PCO camera SDK, National Instrument DAQ driver

Installation steps

  1. Clone this Imagine repository

  2. Install Visual Studio 2015

    • make sure you installed "common tools" when you install
  3. Install QT5

    • make sure the version of QT is for 64bit
  4. Install QT add-in for vc2015

    • install add-in
    • go to QT VS Tools tab in visual studio and set the right directory ex) for vc2015 and Qt5.7 : set this as C:\Qt\5.7\msvc2015_64
  5. Install QWT v6.1.3

    • open the Visual Studio tools directory under the windows Start menu, select "Developer's command prompt"(run as administrator)
    	> cd VC
    	> vcvarsall amd64
    	> cd [qwt directory (extract .zip to c:\ then it will create Qwt-6.1.3 directory)]
    	> C:\Qt_5_7\5.7\msvc2015_64\bin\qmake qwt.pro
    	> nmake (if you recompile everything again, C:\Qwt-6.1.3\src\release directory should be clear)
    	> nmake install
    
    • add an environment variable QWTDIR as C:\qwt-6.1.3
  6. Install Boost v1.57.0

    • extract .zip file to c:\Program Files\boost
    • open the Visual Studio tools directory under the Start
      menu, select "Developer's command prompt"(run as administrator)
        > cd c:\Program Files\boost\boost_1_57_0
        > bootstrap.bat
        > .\b2 address-model=64 toolset=msvc-12.0
      
  7. Install camera SDK

    • run SW_PCOSDKWIN_123

    • Choose interface type ex) Silicon Software Dll MeIV or Camera Link HS

    • Setup target project of visual studio : open project and bring up property page Configuration Properties C/C++>General>Additional>Include Directories

         C:\Users\user\AppData\Roaming\Digital Camera Toolbox\pco.sdk\include
      

      Linker>General>Additional>Library Directories

         C:\Users\user\AppData\Roaming\Digital Camera Toolbox\pco.sdk\lib64
      
         C:\Users\user\AppData\Roaming\Digital Camera Toolbox\pco.sdk\bin64
      

      Linker>Input>Additional>Dependencies

         SC2_cam.lib
      
  8. Install NI(National Instrument) DAQ driver

    • Setup target project of visual studio : open project and bring up property page Configuration Properties C/C++>General>Additional>Include Directories

         C:\Program Files (x86)\National Instruments\NI-DAQ\DAQmx ANSI C Dev\include
      

      Linker>General>Additional>Library Directories

         C:\Program Files (x86)\National Instruments\Shared\ExternalCompilerSupport\C\lib64\msvc
      

      Linker>Input>Additional>Dependencies

         NIDAQmx.lib
      
  9. Modify some items in the preference of Imagine project, if the directory of camera sdk installation is changed from the OCPI's setting.

    • include dir : .....pco.edge\include
    • additional library dir : ...pco.edge\lib64 ...pco.edge\bin64(need this?)
    • additional dependency : ...\lib64\sc2_cam.lib
  10. Install FFTW for frequency analysis of piezo control waveform

    • Download FFTW 3.3.5 Windows : ftp://ftp.fftw.org/pub/fftw/fftw-3.3.5-dll64.zip

    • unzip and copy to c:\FFTW

    • Open the Visual Studio Developer Command prompt Start -> All Apps -> Visual Studio 2015 -> Developer Command prompt Or, run C:\Program Files (x86)\Microsoft Visual Studio 14.0\Common7\Tools\VsDevCmd.bat

    • Produce .lib file : Navigate to the unzip location and type lib /machine:x64 /def:libfftw3-3.def lib /machine:x64 /def:libfftw3f-3.def lib /machine:x64 /def:libfftw3l-3.def

    • Setup visual studio : open project and bring up property page Configuration Properties C/C++>General>Additional>Include Directories

        C:\FFTW
      

      Linker>General>Additional>Library Directories

        C:\FFTW
      

      Linker>Input>Additional>Dependencies

        libfftw3-3.lib, libfftw3f-3.lib, libfftw3l-3.lib
      
  11. Now, we can build Imagine project. But, to execute the Imagine, we need to copy all files below under the 'Output Directory' and 'Working Directory' of Imagine project.

  • Camera related dlls (located in C:\Program Files (x86)\Digital Camera Toolbox\pco.sdk\bin) : SC2_Cam.dll and camera interface dll (sc2_clhs.dll or sc2_cl_me4.dll)
  • QT related dlls (located under the installed directory (ex) C:\Qt\Qt5.7.0\5.7\msvc2015_64\bin) : Qt5Core.dll, Qt5Gui.dll, Qt5OpenGL.dll, Qt5PrintSupport.dll, Qt5Script.dll, Qt5SerialPort.dll, Qt5Svg.dll, Qt5Widgets.dll
  • QWT related dll (located in C:\Qwt-6.1.3\lib): qwt.dll
  • FFTW related dll (located in C:\FFTW): libfftw3-3.dll, libfftw3f-3.dll, libfftw3l-3.dll
  • JavaScript files (located in (Imagine directory)\Imagine\scripts) : imagine.js, 'related rig name'.js ((ex) ocpi-2.js)

Usage

OCPI Operation Modes

  • Parameter control mode
  • Waveform control mode
  • Script control mode
  1. Parameter and waveform control modes

alt text

  1. Script control mode In the script control mode, we can execute javascript code in Imagine. In addition, Imagine provides several javascript functions to provide access to some essential operations which are used to record data. Thereby, we can execute several experiments according to the control flow specified with javascript code by one execution without intervention.

In the Imagine GUI, general settings, parameter control settings and .json file name for waveform control can be saved in a configuration file with a ‘save configuration’ command in file menu. In script control mode, we can load, verify and record with this configuration file.

Typical work flow is as follows.

a. Prepare specimen

b. Capture full live image with a "live" button

c. Set parameters of display : intensity scale, zoom region, etc

d. Set parameters for an experiment : camera, positioner, laser, stimuli, output filename, etc

e. Press "Apply all" button : This verifies if the setting parameters are valid. If not, Imagine rejects the settings

f. Start acquisition : If the parameters are valid, we can start acquisition by clicking a "record" button.

g. Inspect the acquired data : Once the acquisition is end, we can verify the acquisition if it is ended normally or interrupted by some errors, Using "display" tab and "AI and DI" tab, we can verify that and also briefly check if we got the desired data.

Graphic User Interface

alt text

Imagine GUI is composed with several sections as listed below with numbers corresponding the numbers in the above image.

  1. Parameters and display control window

This window is designed for setting overall parameters for the microscope operation and display. This window consists of three main sections which are "Config and control", "Histogram" and "Intensity". Currently, the "Intensity" tab is not used. "Histogram" section displays histogram of an image displayed in "Display window". "Config and control" tab is designed for setting overall parameters such as camera control, positioner control, laser control and stimuli control which are used to configure an experiment. Also, it provides some basic tools for later inspection of acquired data.

  1. Display window

This window displays live images from the cameras attached to the microscope or captured images during the acquisition or after the acquisition by loading the '.imagine' file at the "Display" tab. In this window, we can zoom the image. This zoomed region can be set as a soft ROI region at the "camera" tab.

  1. Log window

This window displays log messages.

  1. Stimuli window

This window makes a list of stimulus numbers according to the stack numbers when we load stimuli file. Details are explained in the explanation of "Stimuli" tab.

  1. Main menu

Details are explained in the explanation of main menu page.

Main Menu

  1. File menu

alt text

We can save a current configuration and load it later using this menu.

  1. Acquisition menu

alt text

We can start acquisition or live displaying the images from the cameras.

  1. Hardware menu

alt text

We can open and close a laser shutter of laser. "Heatsink Fan" and "Temperature" menus are not usable currently.

  1. View menu

alt text

  • Display Full Image : this menu causes the camera to acquire images with the entire sensor.
  • No Auto Scale : does not apply auto scale to the intensity of the captured images in displaying them.
  • Auto Scale On First Frame : applies auto scale to the first frame image of a stack. And following frames are scaled with the scale factor of the first frame of the stack to which they belong.
  • Auto Scale On All Frames : independently auto scales every frame of the captured images.
  • Manual : enables users to set the scale manually. Intensity from "min" value and "max" value will be scaled to 0 to 255. If we overlap the camera1 image with the camera2 image, "cam2 min" and "cam2 max" can be used to set the minimum and maximum intensity value of the camera2 image.
  • Colorize Saturated Pixels : enables display to indicate the saturated pixel with some colors. Blue color pixels indicates pixels which have minus intensity value after scaling. Red is for pixels having intensity over 256.
  • Flicker control : reduces the flicker caused by abnormal black frames between normal frames.

Parameters and display control window

  1. General tab

alt text

This tab is designed for setting the location and filename of output file and adding some comments about the parameter settings of this experiment. These settings and comments can be saved in a configuration file which will be created by clicking the "Save Configuration" item in the "File" menu.

  1. Camera tab

alt text

  • Record both cameras : enables recording from both cameras.
  • number of stacks : sets the number of stacks.
  • number of frames/stack : sets the number of frames in a stack.
  • exposure time(s) : set exposure time. This value is valid only for "External Start" mode in exposure trigger mode.
  • exposure trigger mode : selects exposure trigger mode. Refer to the explanation in the 'trigger mode' section of PCO camera manual.
  • idle time between stacks(s) : sets idle time between stacks.
  • bi-directional imaging : acquires images during positioner "flyback." Refer to the bi-directional imaging waveform below.
  • angle from horizontal(deg) : specifies angle between objective lens and horizontal plane (optional, user info only).
  • objective lens : specifies zoom factor of the objective lens. This is just used as an information.
  • um per pixel (xy plane) : specifies pixel size. This can be also calculated from the zoom factor. This is just used as an information.
  • hor. start : sets the horizontal start of the sensor ROI or displays horizontal start of the soft ROI when we press "use zoom window value" button. Refer to the ROI section.
  • hor. end : sets the horizontal end of the sensor ROI or displays horizontal end of the soft ROI when we press "use zoom window value" button.
  • ver. start : sets the vertical start of the sensor ROI or displays vertical start of the soft ROI when we press "use zoom window value" button.
  • ver. end : sets the vertical end of the sensor ROI or displays vertical end of the soft ROI when we press "use zoom window value" button.
  • use zoom window value : enables soft ROI with the region selected by zoom window. full chip size : enable full size imaging.

Some parameters such as exposure time and idle time between stacks are tightly related with some parameters in positioner tab. And, some parameters related with ROI are restricted according to ROI mode such as soft ROI (details about ROI modes are explained in the PCO camera manual). Imagine internally blocks undesired number of these parameters.

Positioner control waveform and camera control pulse for bi-directional imaging

alt text

  1. Stimuli tab

alt text

  • Stimulus file : selects a .stim file. Then, Imagine will display the specified stimuli description in the editing window and generate stimuli table listed in the Stimuli window.
  • Apply stimuli : applies the stimuli table.

this figure, 'HR_test.stim' file is loaded, which shows an example description of stimuli. Here, several pairs of numbers are listed. The left number is a stack number and the right number is a stimulus number. If "Apply stimuli" is enabled, the stimulus number will be encoded with a 4-bit binary number and then each bits will be used to generate TTL pulses to channel P0.0 ~ P0.3.

  1. Positioner tab

alt text

This tab is for setting the start position and stop position of positioner and also positioner moving back time.

  • Axis : reserved for later use
  • min distance(um) : reserved for later use
  • max distance(um) : reserved for later use
  • start position(um) : sets the start position of the positioner (refer to the positioner control waveform below).
  • move to position(um) : moves positioner to the specified position by clicking the 'move' button.
  • stop position(um) : sets the stop position of the positioner
  • Refresh button : captures the current position of the positioner.
  • time for moving back(s) : Refer to the positioner control waveform below. This time should be longer than (stop position - start position)/(maximum positioner speed) and shorter than idle time between stacks when we don't use bi-directional imaging mode. Imagine can set the shortest time valid automatically by checking the 'Auto' checkbox.
Positioner control waveform and camera control pulse

alt text

  1. Display tab

alt text

  • Display area size : scales the display size.
  • Camera and File choose radio button : chooses which images to display between images from cameras or captured images from .imagine file. And, we can display images from two cameras in one display window by overlapping them with different colors or in two windows separately.
  • Mismatch correction for image 2 : helps to find mismatch parameters between the two overlapped images. When we capture images from two cameras, usually mismatch happens because of the difference between two optical paths to the cameras. We can correct the mismatch between overlapped images in the display window by changing "Translation X","Translation Y" and "Rotation angle" spinbox values. Once a user find well matched parameters with these spinboxes, the user should do the actual correction with those parameters in following analysis with separate tools. Imagine does not do the actual mismatch correction to the data. If this mismatch correction is performed before acquisition, these parameters will be saved in the .imagine output file.
  • Save current image : takes a snapshot of current display window.
  1. Laser tab

alt text

The laser module of OCPI-2 has five different wavelengths. We can turn on and off an individual laser by checking the checkbox beside which the wavelength is labeled. The power of the laser can be also adjusted with slide bars and spinboxes. During the laser setting, we need to open the main laser shutter by clicking the "Open Shutter" button in the main menu. If we use waveform control mode, we can control the each shutters of the five individual wavelengths differently.

  1. Waveform tab

alt text

  • Enable : enables the waveform control mode. If disabled, the parameter control mode is selected.
  • File name :selects a .json custom waveform file.
  • Control info. : brings up pop up window which includes a list of control names and assigned DAQ channels.
  • exp. trigger mode : selects a exp. trigger mode of this waveform control.
  • waveform display controls : enables us to select which waveform and to set a x axis interval or a y interval to display (only display purpose).
  1. AI and DI tab

alt text

  • Imagine file name : selects a .imagine file. Then, Imagine will load .ai and .di file which include data of analog input signals and digital input signals respectively.
  • Waveform display controls : enables us to select which waveform and to set a x axis interval(x min, x max) or a y axis interval(y min, y max) to display. Notice the "Redraw" button. Different from the waveform display control in "waveform" tab, the waveform display window will be updated only when we press this button for the x interval change.
  1. Script tab

alt text

  • Script file : selects a .js script file. Then, Imagine will display the specified script in the editing window
  • Undo : undoes the last change in the editing window.
  • Redo : redoes the last undo in the editing window.
  • Save : saves the script in the editing window to .js file.
  • Execute : executes the script in the editing window.
  • Stop : stops the execution.

In this figure, 'scriptControlTest.js' file is loaded, which shows example script descriptions. This script control mode will be explained in more detail in the next section.

Script control mode

In the script control mode, we can execute javascript code in the Imagine. In addition, Imagine provides several javascript functions for user to access some essential operations which are used to record data. Thereby, we can execute several experiments according to the control flow specified with javascript code by one execution without intervention.

In the Imagine GUI, general settings, parameter control settings and .json file name for waveform control can be saved in an configuration file with a ‘save configuration’ command in file menu. In script control mode, we can load, verify and record with this configuration file.

  1. System functions provided by the Imagine
  • print()

prints string to the log terminal.

ex) print("Hello world!");

  • validityCheck()

applies loaded configuration to data acquisition task and checks the validity of config files. This works as if we press a ‘Apply all’ button in the Imagine GUI after loading configuration files if these files are specified.

ex) var isOK = validityCheck("OCPI_cfg1.txt", "OCPI_cfg2.txt"); var isOK = validityCheck();

  • applyConfiguration()

This works same as validityCheck() function.

ex) applyConfiguration();

  • record()

begins to record. This works as if we press a ‘Record’ button in the Imagine GUI after loading configuration files if these are specifies. If configuration files are not specified, this just records according to the current configuration. This function return ‘false’ if execution time passes ‘timeout’ time, otherwise return ‘true.

ex) var isSucceed = record("OCPI_cfg1.txt", "OCPI_cfg2.txt", timeout); var isSucceed = record(timeout);

  • loadConfig()

loads configuration files.

ex) loadConfig("OCPI_cfg1.txt", "OCPI_cfg2.txt"); loadConfig("OCPI_cfg1.txt“);

  • loadWaveform()

loads a waveform file.

ex) loadWaveform("OCPI_waveform.json");

  • sleepms()

makes the script execution sleep specified times in millisecond.

ex) sleepms(1000); // sleep 1000msec

  • setOutputFilename()

changes output file names with specified file names.

ex) setOutputFilename("t1.imagine","t2.imagine"); setOutputFilename("t1.imagine");

  • getEstimatedRunTime()

returns estimated recording time in second. This time is calculated from total sample number of waveform signal. Therefore, it would take more time in actual acquisition which requires additional preparing and finishing times. These overhead times are less than 30sec in total.

ex) var time = getEstimatedRunTime();

  • stopRecord()

stop recording. This function is useful to get a control again when the Imagine lose a control after record() is fail.

ex) stopRecord();

  • getTimeElapsed(int dt)

return various elapsed times according to 'dt' value.

0: record start to DAQ pulses output start

1: DAQ pulses output start to DAQ pulses output end

2: DAQ pulses output end to recording end

3: current time

ex) total_elapsed_time = getTimeElapsed(0) + getTimeElapsed(1) + getTimeElapsed(2);

ex) function current_time() {return getTimeElapsed(3);}

  1. Javascript objects

We can also use existing javascript objects.

ex) print(Math.sqrt(64)+Math.abs(-1)+Math.log(10));

  1. Example
var ret1, ret2, ret3, ret11, ret22, ret33;

// output filename
var paramCam1ConfigFile = "d:/test/OCPI_cfg1.txt";
var paramCam2ConfigFile = "d:/test/OCPI_cfg1.txt"
var waveCam1ConfigFile = "d:/test/OCPI_cfg1_wav.txt"
var waveCam2ConfigFile = "d:/test/OCPI_cfg1_wav.txt"
var test1Cam1File = "d:/test/test1_cam1.imagine";
var test1Cam2File = "f:/test/test1_cam2.imagine";
var test2Cam1File = "d:/test/test2_cam1.imagine";
var test2Cam2File = "f:/test/test2_cam2.imagine";
var test3Cam1File = "d:/test/test3_cam1.imagine";
var test3Cam2File = "f:/test/test3_cam2.imagine";
var test3WaveformFile = "d:/test/t_script_wav2.json";

// setup recording timeout time
var timeout = 30; // 30sec
var timeoutMargin = 20;

/* Check all the configurations before running them */
print("Checking the validity of all three configurations");
// loard OCPI_cfg1.txt (this is parameter control config file)
loadConfig(paramCam1ConfigFile, paramCam2ConfigFile);
setOutputFilename(test1Cam1File, test1Cam2File);
// check the 1st configuration
ret1 = validityCheck();

// OCPI_cfg1_wav.txt (waveform control : t_script_wav.json is specified in)
loadConfig(waveCam1ConfigFile, waveCam2ConfigFile);
setOutputFilename(test2Cam1File, test2Cam2File);
// check the 2nd configuration
ret2 = validityCheck();

// replace t_script_wav.json with t_script_wav2.json
ret3 = loadWaveform(test3WaveformFile);
setOutputFilename(test3Cam1File, test3Cam2File);
// check the 3rd configuration
ret3 = validityCheck();

/* Execute all the configurations */
if(ret1 && ret2 && ret3) { // If all the configurations are valid

    var estimatedRunTime;
    // setup the 1st configuration
    print("Recoding the 1st configuration");
    loadConfig(paramCam1ConfigFile, paramCam2ConfigFile);
    setOutputFilename(test1Cam1File, test1Cam2File);
    applyConfiguration(); // this can be also used to apply this configuration to the system
    estimatedRunTime = getEstimatedRunTime();
    // Recording
    timeout = estimatedRunTime + timeoutMargin;
    ret11 =  record(timeout);
    if (!ret11) stopRecord();

    // sleep 3000 msec
    print("Waiting 3 seconds");
    sleepms(3000);

    // 2nd configuration
    print("Recoding the 2nd configuration");
    loadConfig(waveCam1ConfigFile, waveCam2ConfigFile);
    setOutputFilename(test2Cam1File, test2Cam2File);
    applyConfiguration();
    estimatedRunTime = getEstimatedRunTime();
    timeout = estimatedRunTime + timeoutMargin;
    ret22 = record(timeout);
    if (!ret22) stopRecord();

    // sleep 100000 msec
    print("Waiting 10 seconds");
    sleepms(10000);

    // 3rd configuration
    // If we are here, we already loaded OCPI_cfg1_wav.txt.
    // So, we just replace waveform file
    print("Recoding the 3rd configuration");
    ret3 = loadWaveform(test3WaveformFile) ;
    setOutputFilename(test3Cam1File, test3Cam2File);
    applyConfiguration();
    estimatedRunTime = getEstimatedRunTime();
    timeout = estimatedRunTime + timeoutMargin;
    ret33 = record(timeout);
    if (!ret33) stopRecord();
}

// Display execution results
print("All the recodings are finished");
var msg1 = "configuration 1 validity: "+ ret1 +", recording: " + ret11;
print(msg1);
var msg2 = "configuration 2 validity: "+ ret2 +", recording: " + ret22;
print(msg2);
var msg3 = "configuration 3 validity: "+ ret3 +", recording: " + ret33;
print(msg3);