Skip to content

Hospira Plum A+ Infusion System version 13.4 and prior,...

Critical severity Unreviewed Published May 13, 2022 to the GitHub Advisory Database • Updated Feb 1, 2023

Package

No package listedSuggest a package

Affected versions

Unknown

Patched versions

Unknown

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

References

Published by the National Vulnerability Database Mar 25, 2019
Published to the GitHub Advisory Database May 13, 2022
Last updated Feb 1, 2023

Severity

Critical
9.8
/ 10

CVSS base metrics

Attack vector
Network
Attack complexity
Low
Privileges required
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Weaknesses

CVE ID

CVE-2015-3954

GHSA ID

GHSA-x5v7-rmgq-cm8v

Source code

No known source code

Dependabot alerts are not supported on this advisory because it does not have a package from a supported ecosystem with an affected and fixed version.

Learn more about GitHub language support

Checking history