Skip to content

Maude specification of a Patient Controlled Analgesia (PCA) open system

Notifications You must be signed in to change notification settings

maude-lang/pca-open-system

Folders and files

NameName
Last commit message
Last commit date

Latest commit

 

History

3 Commits
 
 
 
 
 
 
 
 

Repository files navigation

Programming Open Distributed Systems in Maude

This repository contains a Maude specification and implementation of a Patient Controlled Analgesia (PCA) unit, as a case study for the specification of open distributed systems in Maude. PCA is a widely used method for relieving pain efficiently and safely in hospitalized patients.

The content of this repository is organized in three directories:

  • spec is a self-contained Maude specification of the PCA unit suitable for formal verification. Some safety checks are executed using Maude bounded search. These checks revealed some flaw that let the patient receive an overdose of analgesia, which is fixed in a second version (by replacing the file pca.maude with pca2.maude).
  • impl replaces the external-world components (sensors, actuators, databases, GUI, etc.) of the specification in spec with Maude external objects connecting to external hardware or software outside Maude.

Instructions on how to run these examples can be found in their corresponding directories. Here is a video showing the behavior of the open system implementation.

Open system components

The PCA unit is an object-oriented system that results from the interactions via asynchronous messages of several objects belonging to different classes, each implemented in a separate file:

  • Pca (pca.maude and pca2.maude) is the class of the main object of the PCA specification. It controls and maintains the state of the system, and interacts with all other components.
  • Gui (gui.maude) is in charge of the interaction with the users (patient or clinician) through a graphical user interface.
  • Database (database.maude or mongodb.maude) maintains the logs of the system events for traceability and runtime monitoring using a database.
  • Monitor (monitor.maude) monitors system events through checks on the database to detect undesired behavior and notify the Pca object. In particular, the absence of overdose is checked on a sliding time window.
  • Sensor (sensor.maude) provides on demand measurements of the vitals of the patient (heart rate and temperature).
  • Actuator (actuator.maude) controls the PCA pump that administers analgesia to the patient.

Following the Maude convention, Pca will create most objects dynamically by sending a message to a manager object, guiManager, databaseManager, monitorManager, sensorManager, or actuatorManager. These objects will be fully implemented in Maude in the specification, while they will be or interact with external objects in the implementation. However, the Pca class (as well as the Monitor class) is shared between both variants.

Constants and common declarations for all these objects are kept in open-system.maude. Other files, like configuration.maude, json.maude, or real-time.maude, provide the equational infrastructure for the specification.

This 3 minutes video shows how to run the open system.