Light sheet mode synchronization issue with Photometrics Iris 15 #899

Closed
SJShep opened this issue May 23, 2024 · 2 comments · Fixed by #906
Closed

Light sheet mode synchronization issue with Photometrics Iris 15 #899

SJShep opened this issue May 23, 2024 · 2 comments · Fixed by #906

Comments

@SJShep
Copy link
Contributor

SJShep commented May 23, 2024

Cannot optimize remote focus parameters when using the Photometrics Iris 15 in light sheet mode. There are a few problems, which seem related to synchronization:

  1. Stopping and restarting acquire changes what we see with respect to the remote focusing.
    • image
    • Stop and start acquire: image
  2. After setting the parameters where the ASLM almost looks to be working, changing the exposure causes the current remote focus parameters to fail.
    • image
    • image

Camera Configuration
image

@annie-xd-wang
Copy link
Collaborator

Hi @SJShep, I've made a fix that should solve problem 1. Please test it with the branch #899.

For the waveform parameters, you could try reducing the camera delay time to a shorter value, such as 1 ms. This parameter is set in the configuration.yaml file. At the same time, set an appropriate flyback value and settle-down time for the remote focus through the GUI.

@SJShep
Copy link
Contributor Author

SJShep commented May 24, 2024

That fixed the problem with stopping and starting the acquisitions.