The device shall monitor the battery voltage and trig a high priority alarm Y when autonomy remaining is at 0% (at least 5 min of alarm triggered)
The text was updated successfully, but these errors were encountered:
Update after 30 march meeting: The device shall monitor the battery voltage and trig a medium priority alarm 21 when voltage is at 12.3V
Sorry, something went wrong.
Update after comment on issue #63 : The device shall monitor the battery voltage and trig a High priority alarm 13 when voltage is at 12.0V