Risk Analysis: RCM-SW-12 #59

Closed
Mefl opened this issue Mar 29, 2020 · 2 comments
Closed

Risk Analysis: RCM-SW-12 #59

Mefl opened this issue Mar 29, 2020 · 2 comments
Labels
regulatory Regulatory-related issues software Software-related issues

Comments

@Mefl
Copy link
Contributor

Mefl commented Mar 29, 2020

The device shall monitor the battery voltage and trig a high priority alarm Y when autonomy remaining is at 0% (at least 5 min of alarm triggered)

@cmdm
Copy link

cmdm commented Mar 31, 2020

Update after 30 march meeting:
The device shall monitor the battery voltage and trig a medium priority alarm 21 when voltage is at 12.3V

@valeriansaliou valeriansaliou added regulatory Regulatory-related issues software Software-related issues labels Apr 6, 2020
@valeriansaliou valeriansaliou changed the title Analyse risque : RCM-SW-12 Risk Analysis: RCM-SW-12 Apr 6, 2020
@cmdm
Copy link

cmdm commented Apr 6, 2020

Update after comment on issue #63 :
The device shall monitor the battery voltage and trig a High priority alarm 13 when voltage is at 12.0V