Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

BUG: HS102 Pro Channel 2 AC/DC selection on reconnect #33

Closed
gadjex opened this issue Jun 15, 2020 · 5 comments
Closed

BUG: HS102 Pro Channel 2 AC/DC selection on reconnect #33

gadjex opened this issue Jun 15, 2020 · 5 comments

Comments

@gadjex
Copy link

gadjex commented Jun 15, 2020

On an HS102 Pro, after turning on channel 2 to DC mode and then disconnecting the USB and reconnecting, the app will erroneously show DC mode on channel 2 when it is actually in AC mode. Switching channel 2 off and then back to DC mode will fix the issue until the next disconnect and
reconnect.

This is confirmed by using a volt meter and checking pin PB13 is at 0 volts after reconnecting.

Channel 1 does not display this improper behavior.

@martinloren
Copy link
Owner

Thanks for reporting this issue. What is the firmware version?

@gadjex
Copy link
Author

gadjex commented Jun 19, 2020 via email

@martinloren
Copy link
Owner

martinloren commented Jun 19, 2020

Confirmed, it is a bug on HScope, it will be fixed in next Beta release. By the way there is no documentation for the HS102 Pro but the firmware is working probably....did you build it? I mean with the PGA and all range of input voltages or just single voltage range?

@gadjex
Copy link
Author

gadjex commented Jun 19, 2020 via email

@martinloren
Copy link
Owner

This issue should have been solved in the latest HScope release.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment