Hello,
reset_position will restart a position, from the manager service. You shouldn't need to do this though... it is odd your device reports not having a flowcell - how does it display in the UI at the same time? Are you giving the device enough time to initialise - waiting for a flow cell to be detected? Thanks,
|
Thank you for your response. I have left the device for mins to hours but the message does not change.
however this does not return a response and does not result in a flow cell being detected as I am still receiving the error.
I am using the latest MinION Release 21.11.8. Notably, using an older version (19.12.5) detects the flow cell immediately. |
It sounds like there may be issues with minknow outside of the API, if the UI also doesnt show a flowcell? It may be best to get in contact with support, supply logs for the box you are running minknow on, and a description of the problem, and we can work to solve that issue first. Hope that helps,
|
So I have tried this on another machine (Ubuntu 20.04.3) and again the UI does not detect a flow cell unless I "Restart position". Once the flow cell is detected in the UI, the minknow_api can then detect a flow cell eg. "position_connection.device.get_flow_cell_info()" However, I cannot reproduce "Restart position" of the UI to detect the flow cell using manager.reset_position(). Would this indicate the issue is with minknow_api ? Happy to log a job with support. Very much appreciate your help. |
No, still if the position does not display a flow call shortly after one is connected it seems like a more general hardware or software problem - best to contact support so we can sort that out first. Thanks,
|
Hello ONT,
We are trying to run our minion directly from the command line. For this we are using the start_protocol.py.
We initially had an authentication issue (Issues #34), however we set "--no-tls True" as a workaround.
Now when running we receive the following error:
However, the flow cell is definitely inserted in the minion. I can see that "position_connection.device.get_flow_cell_info()" is used to obtain the flow cell information which is defined in minknow_api.device_service.DeviceService.
Notably, using "position_connection = spec.position.connect()" I can see that the device is running:
Also, using "position_connection.device.get_device_info()" I can return information on the device;
So cool, I can see the device, just not the flow cell. If restart the position using the MiKNOW UI the start_protocol.py script progresses (alas to another issue);
However, I can now retrieve the flow cell info using "position_connection.device.get_flow_cell_info()" after restarting the position with MiKNOW UI
So this leads to my question, how do we restart the position without the MiKNOW UI? I am probably missing something obvious and I apologize in advance. Thank you for any help.
The text was updated successfully, but these errors were encountered: