DS1000Z .wfm files with current firmware #5
Comments
Test.zip VisaDSO shows more sensible values, so you can compare the results. I can make more measurements with different settings, and make screenshots to compare. Feel free to keep/share these files, there is no confidential info in them. |
Thanks Jens! This was very helpful. I fixed a bunch of things and the settings seem correct. (The voltage offsets are currently the negative of what they should be because the conversion for this scope doesn't seem to match that of all the other scopes.) Anyhow, you can look at https://github.com/scottprahl/RigolWFM/blob/master/doc/1-DS1000Z-Waveforms.ipynb to see how things are working. Having all four channels was nice. It seems that the
|
Test.zip |
Thanks again. The calibration pin is a super idea. Hopefully I can get this sorted out today. |
@JensRestemeier I definitely improved the algorithm for calculating voltages for the 1000Z scope, but the offset still does not seem quite right. You can see this the notebooks devoted to the First or Second attempts. So, will you verify that you are using the latest firmware 00.04.04.04.03 from https://www.rigolna.com/firmware/ (it is easy to update, download the firmware to a USB drive, plug it into the scope and agree to update the firmware.) The next thing that might shed some light on the proper algorithm is to capture the same waveform across all four channels with each channel having a different V/div setting and vertical offset. Thanks again for your help |
I downloaded that firmware update on Sunday, though it said something along the lines of "update to same firmware". As the update requires a re-calibration I'd like to avoid that until I have either a specific bug or a specific feature that requires an update. |
so if i wanted to help capture this data for you, what exactly do you need me to do? put 4 probes on calibration point and capture? |
Yes. All four probes to the calibration point would be fine.
|
does it matter if i have used the keygen to unlock all features and 100mhz? |
I don't think so. I really have no idea though. (The goal is to eliminate firmware from the list of issue in converting raw bytes to volts.) |
https://mega.nz/#F!LMVGCKpL!XVG6exOz3_Du0szO5g6ALg |
i dont know if it matters to you but channel 4 is set to 1X mode and the others set to 10X. i figured you might want the variation. |
Got the files. Hopefully this data from a scope with 00.04.04.SP4 firmware will finally make sense. The probe stuff is sort of on the back burner (along with triggering). I don't expect it to confound these results. Thanks again. |
youre welcome and good luck |
NewFolder1.zip |
Thanks Jens. Hopefully I will have time to sort out all the results in the next few days. |
Well, I ran through all contributed 1000Z Thanks again @JensRestemeier and @electronoob. |
The files in the repository are pretty old. The calculated voltages a close to, but do not match the values from the .csv files. The parsing of the header files looks solid and the voltage calculation is the same as that used by all the other Rigol scopes. My hope is that the problem rests in old firmware used to generate the .wfm/.csv files.
The latest scope firmware is at https://www.rigolna.com/firmware/ and a few .wfm/.csv files are needed to test this theory.