CAEN ISPECTOR SDK

Introduction

Ispector use a JSON based REST API for control and configuration of the device. HTTP API allows to perform every operations supported by the web based GUI from a remote host. Ispector use GET/POST HTTP call to both retrieve spectrum/wave data and configure the instruments. HTTP protocol, the same used to transport web page, has the big advantage to easy pass firewall and usually it does not require any special configuration on the network. Several endpoints are available

Endpoint (WEB PAGE) Operation Description
/set_config.cgi POST Set configuration of HV/MCA/PSD
/status.cgi GET Read the status of the instruments and of all available processing channels (MCA/PSD)
/spectrum.cgi GET Get spectrum data
/wavedump.cgi GET Get waveform data
/psd.cgi GET Get
/get_mca_config.cgi GET Readback MCA/PSD configuration
/resetspectrum.cgi GET Reset spectrum
/mca_run.cgi GET Start MCA acquisition
/mca_stop.cgi GET Stop MCA acquisition
/fb_settings.cgi GET Retrieve fabric configuration
/get_sysx.cgi GET Retrieve firmware version and installed options

In order to access to the endpoint perform HTTP GET/POST to address: http://<ispector_ip>/set_config.cgi. Default Ispector IP is: http://192.168.50.2 Ispector API support multiple user simultaneously connected to the instruments. Meanwhile your are developing, keep your browser open on the Ispector page in order see in live the performed operation. Please mind that configuration parameters are loaded on page load. If you change parameters from API you need to refresh browser page in order to see parameters changing on the GUI

Set Configuration

In order to configure the Ispector the /set_config.cgi endpoint is available. This endpoint is a POST service. You need to perform a RAW post POST operation sending in the body of the POST the configuration JSON. Do not use form-data or x-www-form-urlencoded because they are not supported by ISPECTOR There are two different possible configurations you can send to ISPECTOR

  • HV CONFIGURATION
  • MCA CONFIGURATION

HV CONFIGURATION

{"command" : "SET_CHANNEL_CONFIG","channel_config" :[{"id" : 0,"HV_STATUS" : true,"HV_VOLTAGE" : 41.5,"MaxV" : 46,"MaxI" : 5, "RAMP" : 20,"TCoeff" : -34,"HV_MODE" : "temperature","HV_PWRON" : true}],"store_flash" : false} 

All parameters are case sensitive

PARAMETERS VALID VALUES FUNCTION
HV_STATUS true/false Enable/Disable HV
HV_VOLTAGE 22..80 HV voltage Please pay attention to do not destroy the sensor setting too high voltage MaxV 22..80 Max output voltage
MaxI 0..9 Trip Current in mA
RAMP 1..100 Ramp speed of HV
TCoeff -1000 .. 1000 [mV/°C] Temperature compensation coefficient
MODE “digital” “temperature” Enable / Disable temperature compensation on HV
HV_PWRON true/false Power on/off the HV on instrument boot

MCA CONFIGURATION

{"command" : "SET_CHANNEL_CONFIG","mca_config" :[{"id" : 0,"trigger_thrs" : 28,"trigger_inib" : 300,"int_pre" : 300,"int_val" : 10,"int_gain" : 80,"pileup_inib" : 30,"pileup_pen" : 30,"baseline_inib" : 24,"baseline_len" : 256,"taget_run" : 0,"taget_value" : 0,"reset_on_apply" : true}],"store_flash" : false}} 

All parameters are case sensitive

PARAMETERS VALID VALUES FUNCTION
trigger_thrs 10..1000 [int] (LSB) Trigger threshold
trigger_inib 10..1000 [int] (ns) Trigger inhibit after a trigger events. (set in in order to avoid double triggers)
int_pre 0..1000 [int] (ns) Charge integrator pre-trigger integration extension
int_val 0..100 [float] (us) Charge integrator integration time
int_gain 0..1000 [int] Charge integrator GAIN
pileup_inib 0..100 [float] (us) Pileup inhibition after a trigger
pileup_pen 0..100 [float] (us) Pileup penalty if a pileup event occurs
baseline_inib 0..100 [float] (us) Baseline inhibition after a trigger
baseline_len 1024,512,256, 128,64,32,16 Length is samples of the moving average used to calculate the baseline
taget_run 0,1,2 Acquisition run mode 0 – FREE 1 – TIME CONTRAINED (ms) 2 – TOTAL COUNTS ON SPECTRUM taget_value [int] Referring to taget_run parameters, this field specify the run limit. For example to run for 10 seconds set taget_run=1 and taget_value=10000 reset_on_apply true/false Reset spectrum when one or more configuration parameters are changed

Get Instrument Status

Get Instrument Status In order to get the status of the inspector perform GET to the following page /status.cgi

 {
    "command": "GET_SYSTEM_STATUS",
    "Result": "ok",
    "ErrorCode": 0,
    "Reason": "",
    "current_status": {
        "system_status": {
            "temperature": 0,
            "eth_status": 0,
            "eth_ip": "192.168.50.2",
            "last_user_interact": -1,
            "power": "wall",
            "battery": false,
            "battery_life": 0,
            "battery_charge": 0,
            "battery_in_charge": false,
            "remaining_time": 0,
            "battery_voltage": 0,
            "battery_current": 0,
            "battery_temperature": 0,
            "alarm": 0,
            "httpcloud": 0,
            "loracloud": 0
        },
        "channels": [{
            "id": 0,
            "HV_STATUS": true,
            "HV_VOLTAGE": 41.5,
            "HV_MODE": "temperature",
            "COMPL_V": false,
            "COMPL_I": false,
            "Vout": 42.38652,
            "Vref": 1.954313,
            "Iout": 0.3540874,
            "IoutRAW": 0.050250001,
            "Temp": 50.199402,
            "SetPoint": 42.356781,
            "ICR": 1294,
            "OCR": 1254,
            "runtime": 4542,
            "livetime": 4540,
            "sattime": 0,
            "incnt": 5856370,
            "outcnt": 5646006,
            "live": 0.969088,
            "dead": 0.030912,
            "mca_running": 1,
            "mca_status": 0
        }]
    }
}

Get Spectrum

In order to get the status of the inspector perform GET to the following page /spectrum.cgi

{
    "command": "GET_SPECTRUM",
    "Result": "ok",
    "ErrorCode": 0,
    "Reason": "",
    "data": [3, 2, 4, 7, .., 1883]
}

The data field in the JSON contains the spectrum data in a 4096 bin array

Get Waveform

In order to get the status of the inspector perform GET to the following page /wavedump.cg

{
    "command": "GET_WAVEDUMP",
    "Result": "ok",
    "ErrorCode": 0,
    "Reason": "",
    "data": [
        [3421, 0, 0, 0, 1, 0, 0],
        [3425, 0, 0, 0, 1, 0, 0],
        [3425, 0, 0, 0, 1, 0, 0],
        [3423, 0, 0, 0, 1, 0, 0], 
    ]
}

The data field in the JSON contains the waveform data. Each data element is and array of 7 elements:

  1. Analog data
  2. Trigger pulse
  3. Charge Integration Window
  4. PSD Tail integration Window
  5. Baseline restorer status
  6. Pile up rejector discard
  7. Pile up inhibition

Readback MCA configuration

In order to get the status of the inspector perform GET to the following page /get_mca_config.cgi

Testing API

Use CURL to POST/GET data EXAMPLE: Send configuration for MCA

	curl -d '{"command" : "SET_CHANNEL_CONFIG","mca_config" :[{"id" : 0,"trigger_thrs" : 100,"trigger_inib" : 100,"int_pre" : 300,"int_val" : 10,"int_gain" : 100,"pileup_inib" : 10,"pileup_pen" : 10,"baseline_inib" : 10,"rebinnig" : 4096,"baseline_len" : 512,"taget_run" : 0,"taget_value" : 0,"reset_on_apply" : true}],"store_flash" : false}}' -H "Content-Type: application/json" -X  POST http://192.168.50.2/set_config.cgi 

EXAMPLE: Read system status curl -X POST http://192.168.50.2/status.cgi

EXAMPLE: Download spectrum curl -X POST http://192.168.50.2/spectrum.cgi

Python SDK

An SDK for Python language is available. The SDK use HTTP API communication to interface with the Ispector. The SDK requires Python >3.2 and works both on x86/ia64/ARM processor. It could be used as well as on a standard PC and on a Raspberry PI.

REQUIRED MODULES FOR SDK:

The SDK requires the following module:

  • requests [pip install requests]
  • enum [pip install enum]

The Example file requires the following module:

  • pprint [pip install pprint]
  • numpy [pip install numpy]
  • matplotlib [pip install matplotlib]

DOWNLOAD THE SDK:

Python SDK files can be download from Nuclear Instruments Github https://github.com/NuclearInstruments/IspectorSDK-Python or cloned with git: git clone https://github.com/NuclearInstruments/IspectorSDK-Python.git

The SDK include library (inspector_sdk.py) and an example file (test_ispector_sdk.py)

LIBRARY USAGE

In order to use the library, import ispector_sdk and open a connection creating a new ispector_sdk object

from  ispector_sdk  import  ispector_sdk
I1  =  ispector_sdk("192.168.50.2")

Have a look to example file and pdf in docs to further informations