Repository navigation
v2.0.2 - XLE voltage scale, chest leads and recording date
ECGDataKit 2.0.2
Updates to the experimental XLE parser.
- Samples are now converted to mV with a scale of 0.9375 µV per count. The file does not store this scale and the manufacturer has not validated it, so the conversion issues a warning.
- Leads are named I, II and CR1 to CR6, since the chest leads are stored against the right arm. An optional Wilson correction,
XLEParser(wilson_correction=True), turns them into V1 to V6, with a warning. - The recording date and time are now read from the file.
XLE support remains experimental. Check amplitudes against the recorder's own printout before use.
Full Changelog: v2.0.1...v2.0.2