ECGDataKit 2.0.2

Updates to the experimental XLE parser.

  • Samples are now converted to mV with a scale of 0.9375 µV per count. The file does not store this scale and the manufacturer has not validated it, so the conversion issues a warning.
  • Leads are named I, II and CR1 to CR6, since the chest leads are stored against the right arm. An optional Wilson correction, XLEParser(wilson_correction=True), turns them into V1 to V6, with a warning.
  • The recording date and time are now read from the file.

XLE support remains experimental. Check amplitudes against the recorder's own printout before use.