Skip to content

First Steps

Julian Decker edited this page Nov 12, 2026 · 12 revisions

First steps

The first start

You see the Rigol logo from the stock loader, then the OOS start-up screen with a checklist, then the scope. Each failed step is named on the screen. The version and the build stand at the bottom of the start-up screen. A start takes about 8 seconds.

The IP address comes from DHCP and is shown at the top right. Without DHCP the scope takes a link-local address, 169.254.x.y, as computers do (from 0.1.9; before, 192.168.177.64/24): a computer connected straight to the scope reaches it without any set-up.

OOS starts in one-shot mode: the next power-on starts the stock system. When OOS works for you, set System > Boot mode to Persist. Switch the scope off at the mains and wait 30 seconds between two starts of OOS.

The screen

Main screen

  • Top bar: run state, sample rate and memory depth, the menus, network and USB.
  • Plot: the waveforms. Drag offset, delay and trigger level, pinch to zoom. The button with four corners at its top right gives the plot the whole screen (below).
  • Strip: the measurements.
  • Cards: Menu, channels, logic analyser, generator, time base and trigger. Tap a card for its settings.

Front-panel keys and knobs work as labelled. The Menu button reaches every function. From 0.1.9 a channel key held for more than half a second, or a finger held on a channel's card, switches that channel between AC and DC coupling; channel keys act when they are let go.

Sample rate

Tap the sample rate at the top for Acquire. The rate itself follows the time base and the memory depth: slower time base or less memory, lower rate. The top rate is 1.25 GSa/s with one channel, 625 MSa/s with two, 312.5 MSa/s with three or four.

Auto

The Auto key sets up the channels that are on: levels, time base and trigger, and it sets the memory depth back to Auto. It searches at full bandwidth and then gives each channel its own bandwidth back. With no signal it restores your set-up.

After Auto

After Auto a bar at the top of the plot offers four views of the trigger channel:

Button View
a few pulses about three periods
many pulses about ten periods
rising ramp the rising edge across about two divisions, trigger on the rising edge
falling ramp the falling edge, trigger on the falling edge

The x or a tap anywhere else closes the bar; otherwise it goes after 30 seconds.

Windows and full screen

Display > Windows splits the plot:

Layout Windows
One the whole plot
Side by side two, left and right
Stacked two, one above the other
Four two by two

On the same page each channel, each math channel that is on and the logic analyser get a window. A new layout spreads the channels that are on, one per window, and you can move them afterwards. All windows share the time base. Each window has its own channel markers, volt labels and time axis. A drag in a window moves a channel of that window. The logic analyser lanes stand at the bottom of their window; with no channel in that window they fill it. Zoom, XY and the analysers use one window. The layout is saved with the setup.

Full screen

Full screen: the button with four corners at the top right of the plot hides the top bar and the cards. The same button, now with the corners pointing inwards, brings them back. Also on Display > Windows > Full screen. While the touch lock is on there is no full screen, because the Menu button is the way out of the lock.

Bandwidth and noisy signals

A channel's bandwidth is Auto unless you choose another: the limit follows the sample rate, as on the stock system, so a slow time base shows less noise. With Full, a signal well above that limit keeps its size, but the noise of a slow signal can be larger than the trigger's hysteresis, and the hardware then triggers on the other edge too. The screen still shows the edge you set. Trigger > Noise reject widens the hysteresis from 0.3 to 0.7 divisions.

Text size

From 0.1.6. Display > Text size: Normal, Large or Larger. Large and Larger make the text of the bars, cards, measurement rows, menus and dialogs 2 or 4 pixels bigger; the plot gets a little shorter. Kept across restarts.

Display settings

  • Display > Backlight (from 0.1.9): the panel's brightness, 10 to 100 %, in steps that look even.
  • Display > Graphics (from 0.1.7): GPU (the default) or CPU. With GPU the scope's graphics chip draws the picture, which takes about a third of the load off the processor; from 0.1.8 on the chip is switched on. Where it does not start, the CPU draws.
  • Display > HDMI (from 0.1.7): Mirror, Waveform, or Layout, which lays the whole screen out again at the monitor's size.

The keyboard

Names of channels and references, math formulas and WLAN passwords are typed on the on-screen keyboard. The arrows either side of Space move the insertion point one character, and a tap into the entry puts it where you tap. Typing and Delete work at the insertion point, so a wrong character in a long password is fixed without typing the rest again. Show / Hide shows the password.

Saving

Setups stay on the scope. Screenshots and CSV files go to a USB stick in the front port, into the folder OOS.

Date and time

From 0.1.6. The scope has no clock battery and starts in 2013. System > Date and time:

  • Time from NTP (off by default; then OOS sends nothing to the outside): the clock from pool.ntp.org or another server, a few seconds after a network link comes up and then every hour.
  • Time zone: common zones with their daylight saving time, or any POSIX TZ string.
  • Set the clock: date and time by hand, for a scope without a network.

Once the clock is set, new files are named by date and time (shot-20261111-143005.png) instead of a number.

Self-calibration

System > Self-cal. Warm the scope up for 30 minutes and disconnect all inputs. It takes about 5 minutes per channel and pass, with 4 passes. The factory files are never changed.

Clone this wiki locally