GUI Guide
Control Tab
- Choose USB port dedicated for ESP32. Auto-detection will pick the most probable port.
- If the port does not show up, look for the USB port by following multiple USB port guide
- When the disconnection button is clicked, all channels will be set to
IDLEmode before full disconnection.
![]()
After connecting, always reboot ESP32 to avoid initialization failure
Initialization failure occurs when USB is connected before 9V power adapter is plugged in
- Source steady biasing current at set voltage
-
$1 k \Omega$ and$100 k \Omega$ resistors are in series to supply current to the nano-wire with set DAC voltage - Current read-back on the GUI allows easy current measurement
Required User Inputs:
- Voltage in V
- (Optional) Timer (Time units: Min, Hour, Day, Month)
- When the timer is set, the specified channel will supply the set voltage for the set duration
- When electrometer is connected in
calib taband timer is set, comparison plot will be available as a clickable link under STOP button or as a txt file inoutput/directory after timer is done - Electrometer comparison plot is useful for comparing calibrated current sensing on PCB to a commercial instruments for reliability of the PCB measurement
- Hysteresis V-I sweep at set voltage range and voltage step size
- Quicker sweep can be done by either changing to higher ODR, bigger step size, or smaller range
- However, there is tradeoff between time and precision
Required User Inputs:
- Voltage Range (Default:
$\pm 0.5V$ , software cap at$\pm 1.5 V$ )- Step Size (Default:
$10 m V$ )
For v1.0 PCB
PCB Component Limit: Maximum sweep range is $\pm 1.5 V$
- Op-amp's reference voltage is
$\pm 2.5 V$ - Maximum positive common mode input voltage that can maintain linearity is
$V_{REF+} - 1.5 V = 1.0 V$ Reference voltage will be altered to
$5.0V$ for v2.0 PCB
- Sigma-Delta ADC over-samples and averages out samples to reduce noise
- The lower the Sample per Second (SPS), the less the noise but at the cost of integration time
- Default is set to 10 SPS, which is slightly faster than commercial electrometer's default NPLC settings, but with slightly less precision
- User can change this ODR setting from the GUI from the options in the figure above
Channel ID
- Channels on the GUI correspond to the PCB channels as in the above image
The more channel enabled (Not in OFF mode), the slower the measurement
Calibration Tab
- User can choose the number of electrometer up to 4 devices
- Check the USB TMC connection for each device
- Choose channel for steady current electrometer comparison
- Sets the polling rate for each electrometer
- The steady comparison data between PCB ADC and electrometer will be collected once every polling rate
- Calibrates components that require manual electrometer measurement
- The resistors and pins specified in the image below need to be calibrated with precision electrometer (e.g. DMM 6500 6.5 digit multimeter)
![]() |
![]() |
|---|---|
| Required resistors and pin for calibration | Probe DAC's V_REFIN pin (4th pin) for 2.5V |
-
Set electrometer NPLC to its maximum
-
4-wire Kelvin Measurement on
$402 \Omega$ &$1 k \Omega$ resistors -
Measure 2.5V DAC reference voltage
-
Input measured values on the GUI in
Calib tab -
Apply values by clicking
calibrate -
Repeat step 1 ~ 5 for the rest of the channels
Required User Inputs:
$R_{sense} = 1 k \Omega \pm 1%$ ,$R_{gain} = 402 \Omega \pm 1%$ ,$V_{REF_{DAC}} = 2.5 V \pm 0.5%$
- If measured value is out of the tolerance, GUI will show an error. Measure again and ultimately swap the resistors
Non-Volatile Memory (NVM)
- Data stored in NVM last across power cycle (e.g. reboot, power loss, reflashing firmware)
--> Previous calibrated data can be loaded on every boot - Fresh firmware will have ideal resistor and reference voltage values, so calibration is not strictly necessary
- However, it is strongly recommended to do 4-wire Kelvin measurement calibration on every channel for precise current sensing
- Clicking
fetchwill fetch current calibrated values from the ESP32