Skip to content

Initial Calibration

ta1_k edited this page Aug 24, 2026 · 5 revisions

Calibrate current sensing resistors & DAC reference voltage

Required resistors and pin for calibration Probe DAC's V_REFIN pin (4th pin) for 2.5V
  1. Set electrometer NPLC to its maximum

  2. 4-wire Kelvin Measurement on $402 \Omega$ & $1 k \Omega$ resistors

  3. Measure 2.5V DAC reference voltage

  4. Input measured values on the GUI in Calib tab

  5. Apply values by clicking calibrate

  6. Repeat step 1 ~ 5 for the rest of the channels

Clone this wiki locally