A new workspace: navigation with the runs on the left, reference and control chips on top, a hint line that says what is missing next, a start screen with drag and drop and a demo project.
Curve analysis: baseline, Cq at a common threshold, per-well efficiency in a window of linearity, melt peaks and Tm, compared with instrument results on real files. The instrument's Cq stays the default; Cq and efficiency from curves are an explicit choice.
The curve graph beside the plate and linked with it, a threshold line that can be dragged, per-run manual thresholds with an Auto button.
New QC from curves: several melt peaks, Tm shift, a product or dimers in the NTC, disagreement with the instrument's Cq, very early Cq, Cq without amplification, mixed Cq sources, thresholds that differ without calibrators.
Exclude a standard by clicking its point; save the expression figure from its page; curve columns in the data table, a Curves sheet in the workbook, curve figures in the HTML report and in the exported figures; qpcrkit analyze --cq-source.
QuantStudio Design & Analysis workbooks; RDML 1.0 and 1.1; several CFX files or a whole folder as one run.
Amplification and melt curves are imported and stored in the project.
Assign targets to wells, rename a target on a plate, name samples in series, paste a plate map from a spreadsheet.
No-RT control role (nrt) with its own check.
Reference genes and calibrator named by the file are offered after import.
Changed
A calm theme with a teal accent and line icons; analysis settings moved to a dialog, reference and control to the top bar.
Project files are smaller: curves of a run share one axis.
The theme follows the system by default; light and dark can still be chosen.
Wells without a sample are no longer treated as replicates of each other.
Fixed
Checked boxes show a tick; lists and spin boxes show their arrows.
Numbers are read by column: 1,250.000 and 100,000,000 no longer lose standard quantities, and a Russian dilution series keeps its decimals.
Binary instrument files give a clear message.
QC messages show the thresholds from the settings.
Command line: a missing file, a locked xlsx or an unknown --ref/--control give a message instead of a traceback.
Target names cannot break figure file names; the standards tab keeps the selected curve.
A wrong quantity is refused instead of breaking the standard curve.