Repository navigation
OOS 0.1.11
A release of OOS, an open operating system for the Rigol DHO800/DHO900: the HDMI flicker with GPU drawing, frozen measurements, deep memory at short time bases, and from the forum thread the AC trigger and the Bode plot with a generator on the USB port.
Experimental. It has run on a DHO924S, and DHO804s run earlier versions. Read the disclaimer, the status and the known limits first. Measurements are not calibrated to any standard.
Installing and updating
No build needed: Installation.
- From 0.1.7 or newer: install
oos-0.1.11.oosupdthroughhttp://<scope>:8080/update(or System > Update). The scope restarts into the new build by itself (kexec) and installs it from there. Then switch it off at the mains, wait 30 seconds and switch it on: that start is the trial. - From 0.1.6 or older: install 0.1.7 first, the same way, then this one (details in the 0.1.9 release).
- USB stick (
DHO800_DHO900_OOS_Install.GEL, Storage > Upgrade in the stock system): installs any version directly, and also updates an older OOS build on the card.
What changed
New in 0.1.11
- Trigger > Coupling for the edge trigger: DC, AC, LF reject, HF reject (
:TRIGger:COUPling). AC and LF reject take the DC out of the trigger path, so a level of 0 V is the zero crossing of the signal whatever its offset; HF reject keeps the DC and suppresses fast signals. The trace itself is not filtered. - Bode with an external generator on the USB-A port (USBTMC, no PC): Analysers > Bode > Connection = USB takes the first USBTMC instrument and shows its
*IDN?answer or why there is none. New preset for the UNI-T UTG900 (UTG932, UTG962), commands from UNI-T's programming manual; Rigol DG and Siglent SDG work over USB as well. Tested with a simulated device only: reports with a real generator are welcome.
Fixed in 0.1.11
- HDMI with Display > Graphics = GPU: the panel flashed black now and then, mostly with a menu open, and the picture could be broken on both screens. The panel's third display buffer and the monitor's first one shared a place in the GPU's list, and the monitor's frame overwrote the panel's while the GPU still drew it. On a DHO924S with a 1920x1080 monitor, panel and monitor are steady in Mirror, Layout and Waveform.
- Measurements froze in Run (since 0.1.7): values, statistics, the trend and the counter from the record only changed after STOP. The measurements run beside the frame loop, and a result was dropped whenever a newer record had come, which in Run is nearly every frame.
- Deep memory at short time bases: above 1 M points the screen got the whole record compressed to about 625 k min/max points, so at 10 M and 100 ns/div about 80 of them for the screen: a sine showed as a staircase, and only about 3 records came per second. Such a record is now read for the screen as a window at every sample: a clean sine at 10 M and 50 M, 21 to 27 records per second.
- NORMAL without a trigger: a changed trigger took up to 2 s to act; it now acts at once.
- Web remote: a PNG screenshot asked for while the live view ran as JPEG could come back as JPEG bytes; one error of the video encoder no longer switches JPEG off until a restart.
- Display: a late page flip event no longer switches the panel to a slower path for the rest of the session.
Full list: CHANGELOG.md.
Files
| File | What |
|---|---|
oos-0.1.11.img |
boot image, for the second-card install in the README |
oos-0.1.11.oosupd |
update package for a running OOS (0.1.7 or later), signed with the maintainer key (id 4dbd03d9763869dd), which every build trusts |
DHO800_DHO900_OOS_Install.GEL |
install from a USB stick through the stock system's Storage > Upgrade, without opening the scope; keep its name |
SHA256SUMS |
checksums |
linux-6.12.115-base.tar.xz |
kernel source, needed to build the kernel: put it into kernel/src/ |
busybox_1.37.0*, e2fsprogs-1.47.2.tar.xz, wpa_supplicant-2.11.tar.gz, kexec-tools-2.0.32.tar.xz |
sources of the other programs in the image |
mesa_26.0.8.orig.tar.xz, glibc_2.43*, gcc-16_16-20260322*, libdrm_2.4.131*, zlib_1.3* |
sources of the GPU helper's libraries (Mesa, glibc, libstdc++ and libgcc, libdrm, zlib) |
State
On a DHO924S: this exact package installed from a running OOS through the web page, started by kexec and installed; the device test of the quick tier passed on it (15 of 15). The fixes were measured on the unit before (full test rounds, HDMI with a 1920x1080 monitor in all three modes). Bode over USB is tested with a simulated device only.
Bug reports, test results from other models and code are welcome.