MSpektra 5.31

Choose a tag to compare

@its-kietsu its-kietsu released this 18 Nov 10:08

MSpektra 5.31

A small Windows update with calibration fixes for HRMS Analysis.

Calibration (HRMS)

  • HPC now works on the flight time, as in Bruker DataAnalysis. It matches much better and gives the same result every time.
  • A calibrant peak at the edge of its search range is now measured at its top.
  • Saved analyses with a polynomial calibration open again.
  • The calibration you set is kept when you switch between files.
  • The calibrant marker, legend and range stay with their own file when you switch between open files.

Updating

  • From 4.26 or newer: click Update in the bar. Your settings and saved analyses stay as they are.
  • The previous version can be restored from About.
  • From 4.25 or older: download MSpektra-5.31.zip and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.31.zip below.

MSpektra 5.20

Choose a tag to compare

@its-kietsu its-kietsu released this 16 Nov 07:19

MSpektra 5.20

A Windows update: one window for everything, a new file list, other LC detectors, and many smaller improvements.

One window

  • LC Analysis, HRMS Analysis and Deconvolute now open in the same window.
  • A Home button (top left) takes you back. Alt+Home does the same.
  • Home shows what is open now, your recent files and new data in your folders.
  • MSpektra starts faster.

LC Analysis (was LCMS Analysis)

  • The new name fits HPLC data as well as LC-MS. Reports are now called "LC purity report".
  • Other detectors: RI, CAD, ELSD, fluorescence and other channels each get their own view, with the MS spectra at the selected time below, like the PDA view. Choose the detector with the button next to PDA (for example "RI ▾"). Read from Waters, Agilent and Shimadzu files.
  • PDA: right-click the chromatogram to add or remove a wavelength, and to show the wavelengths overlaid or stacked.
  • The mass spectrum opens with the m/z range the file records, also after averaging.

File list

  • Top half: your pinned and recent folders with their newest files. One click opens a file. Folders show their full path; subfolders open inside them.
  • Bottom half: OPEN NOW, the files you have open. Drag its heading to change the size.
  • Right-click a recent folder to remove it from the list.

Reports

  • A clean white look in PDF and Word, with tables as wide as the plots.

Help and safety

  • Send feedback: on Home, just above Help.
  • If MSpektra closes unexpectedly, the next start offers to reopen your files and to send a short report.
  • Help > Reset settings… puts the settings back to how they were at first start. Your data, saved analyses and saved methods are not changed.

Compare

  • Pan moves along the x axis only and can go beyond the range you set.
  • The mouse wheel makes the peaks taller or lower. Double click or Fit sets it back.
  • The Select tool zooms in with a drag.

Fixes

  • Opening LC or HRMS no longer flashes two small windows.
  • MSpektra closes cleanly.

Updating

  • From 4.26 or newer: click Update in the bar. Your settings and saved analyses stay as they are.
  • The previous version can be restored from About.
  • From 4.25 or older: download MSpektra-5.20.zip and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.20.zip below.

MSpektra 5.18

Choose a tag to compare

@its-kietsu its-kietsu released this 12 Nov 17:18

MSpektra 5.18

A Windows update: clearer Compare plots and images.

  • Zooming in on stacked spectra keeps every row whole.
  • Spectrum names sit inside each plot, at the top right, so the plot gets more room. "Right of the frame" is still under Display > Labels.
  • New scale: Relative: tallest peak in view = 100 %. Each spectrum fits the height of what you zoomed to, as in the MS view.
  • The scale names are clearer: Absolute, Relative: tallest peak of each = 100 %, Relative: peak at the reference time = 100 %.
  • Copied images (for Word and other programs) show the title at the top left, above the plot.

Updating

  • From 4.26 or newer: click Update in the bar. Your settings and saved analyses stay as they are.
  • The previous version can be restored from About.
  • From 4.25 or older: download MSpektra-5.18.zip and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.18.zip below.

Built from its-kietsu/MSpektra-cpp commit a71216afe41510aa49bfbc43cb3598171fa2e5ed.

MSpektra 5.17

Choose a tag to compare

@its-kietsu its-kietsu released this 12 Nov 01:56

MSpektra 5.17

A Windows update: Compare no longer stops the window, and goes from the chromatogram to the spectra and the masses.

  • Compare no longer stops responding. The mass spectra, the masses and HRMS mass chromatograms are made in the background, and large spectra draw quickly.
  • Mass spectrum without a time range: each run's main peak, not the whole run.
  • Drag on the plot to zoom in. Fit or Home shows everything again.
  • Right click a chromatogram: Mass spectra of this time range.
  • Right click the spectra: Deconvolve these spectra. The masses use your deconvolution settings.
  • Right click a mass: Mark its charge states on the spectra. You can also type a mass under Signal.
  • The peaks of spectra and masses are labelled as in the MS view.
  • A new calibration of an HRMS file updates its spectra and masses in Compare. The table shows each file's calibration.
  • The Templates window opens at once.

Updating

  • From 4.26 or newer: click Update in the bar. Your settings and saved analyses stay as they are.
  • The previous version can be restored from About.
  • From 4.25 or older: download MSpektra-5.17.zip and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.17.zip below.

Built from its-kietsu/MSpektra-cpp commit ef2f6145a2f80b608b477f36573ad4ef502cb2b6.

MSpektra 5.16

Choose a tag to compare

@its-kietsu its-kietsu released this 11 Nov 21:13

MSpektra 5.16

A Windows update: Compare and PDA views in HRMS Analysis, new Compare styles, and hybrid deconvolution no longer crashes.

  • Hybrid deconvolution no longer crashes on LCMS spectra with an m/z range and several charges.
  • The same fix in Bayesian deconvolution. Results are unchanged, except with the split Gaussian peak shape: up to 0.2 % of the top.
  • Deconvolution results: right click > Delete all mass labels. Show the mass labels again brings them back.
  • HRMS Analysis has three views like LCMS Analysis: MS, PDA (UV/Vis) and Compare. UV data of Thermo and Agilent files is read.
  • Compare: compare the TIC, BPC, an extracted ion chromatogram (EIC), the UV chromatogram or the averaged mass spectrum.
  • Compare: new styles for the layout: Framed panels, Joined panels, Banded rows, Ridgeline, Filled overlay and Mirror. Templates shows them all with your traces.
  • Find mass finds much weaker proteins, also large ones and ones with few charge states.
  • Saving no longer stops the window, also with many files and results, and it is several times faster.
  • Closing files gives back the memory of their results.
  • The program, the help and the plots use the Segoe UI Variable font (Segoe UI on Windows 10).
  • Copied and saved plots show the scan, time and base peak at the top right. Peak labels stay clear of it.
  • LCMS Analysis: added plot tiles are as tall as the chromatogram.
  • HRMS Analysis: clicking a deconvolution result no longer scrolls the window. A double click resets its zoom in place.
  • The Rename Files window shows its full title.

Updating

  • From 4.26 or newer: click Update in the bar. Your settings and saved analyses stay as they are.
  • The previous version can be restored from About.
  • From 4.25 or older: download MSpektra-5.16.zip and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.16.zip below.

Built from its-kietsu/MSpektra-cpp commit 84ffb1402377c5de56636fa4e11e2f7f97569af0.

MSpektra 5.15

Choose a tag to compare

@its-kietsu its-kietsu released this 10 Nov 11:16

MSpektra 5.15

A Windows update with fixes to deconvolution labels, Find mass and saved analyses.

  • Deconvolution peak labels: right click a peak to remove its label, including an automatic label. Choose Label to restore it. Hidden labels stay hidden when the selection or zoom changes and when an analysis is reopened.
  • Removing the label of a manually added mass keeps the mass in the table. Removing that mass remains a separate action.
  • LCMS Analysis: opening more chromatogram, mass search or deconvolution tiles keeps existing plots at readable heights. Scroll down for the added tiles.
  • Find mass: a stronger species outside the requested mass window no longer rejects a target inside it.
  • Formula finder: changing the search inputs clears old candidates so they cannot be shown with the wrong ion.
  • Polymer series keep the first isotope mass after a charge edit. Revised series update their mass distributions.
  • A failed update rollback keeps its recovery files and can be retried.

The unfinished hybrid crash fix is not included in this release.

Updating

  • From 4.26 or newer: click Update in the bar. Your settings and saved analyses stay as they are.
  • The previous version can be restored from About.
  • From 4.25 or older: download MSpektra-5.15.zip and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.15.zip below.

Built from its-kietsu/MSpektra-cpp commit a2e2dd3e116a44f749f7b7d8150a83246d11a604.

MSpektra 5.14

Choose a tag to compare

@its-kietsu its-kietsu released this 05 Nov 08:25

MSpektra 5.14

A new, much faster MSpektra. The windows, tools and files are the same as in 4.59.

New in 5.14

  • Peak areas are measured from the real baseline, not a straight line between the ends of the peak. Areas differ from earlier versions.
  • Peaks in a saved analysis keep their old area until you integrate them again.
  • The injection at the start of a run is no longer the main peak. It is left out of the Area % and the purity.
  • MS spectra button in PDA: shows or hides the mass spectra.
  • The Method button opens its menu at the button, and the Method title at the title.
  • Kinetic fitting: a Time points window.
  • Clicking a saved analysis opens only that file.

New in 5.13

  • The side panel has tabs: one page at a time, less to look through.
  • The scroll bar of the side panel can be seen and dragged.
  • Tables: drag the border of a column to change its width. A double click fits it again.
  • Compare table: only what you need on screen. Copy table, export and report keep everything.
  • Choosing another file keeps you on the same view (MS, PDA or Compare).
  • The side panel left open keeps all its tabs at the next start.

New in 5.12

  • A saved analysis now opens in the same window, next to the files already open.
  • File > Preferences: choose one folder for your saved analyses. Left empty, they stay next to the data files.
  • The Calibration window opens smaller.
  • Calibration: the masses on the spectrum no longer touch the axis, and the result note always shows in full.

New in 5.11

  • A peak under the description at the top of a spectrum keeps its label. The label moves just below the description.

New in 5.1

  • Peak labels: the Label tool and the right-click menu remove any label, also the automatic ones.
  • Display > Peak labels turns the automatic labels on or off.
  • A tool button clicked again turns the tool off.

New in 5.0

  • Much faster: it starts, opens files and draws plots in a fraction of the time.
  • Much smaller: the download is 33 MB (was 309 MB), the folder 82 MB (was about 870 MB).
  • Bigger plots: thin title strips and small gaps give the plots the space; no label is cut.
  • After a deconvolution the chromatogram, the spectrum and the result fit the window; the mass table is below.
  • A tidy folder: MSpektra.exe, README, LICENSES and the folder app.

Updating

  • From 4.26 or newer (also 5.0, 5.1, 5.11, 5.12 and 5.13): click Update in the bar. Your settings and saved analyses stay as they are.
  • The version before can be put back from About.
  • From 4.25 or older: download MSpektra-5.14.zip below and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.14.zip below.

MSpektra 5.13

Choose a tag to compare

@its-kietsu its-kietsu released this 04 Nov 08:39

MSpektra 5.13

A new, much faster MSpektra. The windows, tools and files are the same as in 4.59.

New in 5.13

  • The side panel has tabs: one page at a time, less to look through.
  • The scroll bar of the side panel can be seen and dragged.
  • Tables: drag the border of a column to change its width. A double click fits it again.
  • Compare table: only what you need on screen. Copy table, export and report keep everything.
  • Choosing another file keeps you on the same view (MS, PDA or Compare).
  • The side panel left open keeps all its tabs at the next start.

New in 5.12

  • A saved analysis now opens in the same window, next to the files already open.
  • File > Preferences: choose one folder for your saved analyses. Left empty, they stay next to the data files.
  • The Calibration window opens smaller.
  • Calibration: the masses on the spectrum no longer touch the axis, and the result note always shows in full.

New in 5.11

  • A peak under the description at the top of a spectrum keeps its label. The label moves just below the description.

New in 5.1

  • Peak labels: the Label tool and the right-click menu remove any label, also the automatic ones.
  • Display > Peak labels turns the automatic labels on or off.
  • A tool button clicked again turns the tool off.

New in 5.0

  • Much faster: it starts, opens files and draws plots in a fraction of the time.
  • Much smaller: the download is 33 MB (was 309 MB), the folder 82 MB (was about 870 MB).
  • Bigger plots: thin title strips and small gaps give the plots the space; no label is cut.
  • After a deconvolution the chromatogram, the spectrum and the result fit the window; the mass table is below.
  • A tidy folder: MSpektra.exe, README, LICENSES and the folder app.

Updating

  • From 4.26 or newer (also 5.0, 5.1, 5.11 and 5.12): click Update in the bar. Your settings and saved analyses stay as they are.
  • The version before can be put back from About.
  • From 4.25 or older: download MSpektra-5.13.zip below and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.13.zip below.

MSpektra 5.12

Choose a tag to compare

@its-kietsu its-kietsu released this 02 Nov 14:57

MSpektra 5.12

A new, much faster MSpektra. The windows, tools and files are the same as in 4.59.

New in 5.12

  • A saved analysis now opens in the same window, next to the files already open.
  • File > Preferences: choose one folder for your saved analyses. Left empty, they stay next to the data files.
  • The Calibration window opens smaller.
  • Calibration: the masses on the spectrum no longer touch the axis, and the result note always shows in full.

New in 5.11

  • A peak under the description at the top of a spectrum keeps its label. The label moves just below the description.

New in 5.1

  • Peak labels: the Label tool and the right-click menu remove any label, also the automatic ones.
  • Display > Peak labels turns the automatic labels on or off.
  • A tool button clicked again turns the tool off.

New in 5.0

  • Much faster: it starts, opens files and draws plots in a fraction of the time.
  • Much smaller: the download is 33 MB (was 309 MB), the folder 82 MB (was about 870 MB).
  • Bigger plots: thin title strips and small gaps give the plots the space; no label is cut.
  • After a deconvolution the chromatogram, the spectrum and the result fit the window; the mass table is below.
  • A tidy folder: MSpektra.exe, README, LICENSES and the folder app.

Updating

  • From 4.26 or newer (also 5.0, 5.1 and 5.11): click Update in the bar. Your settings and saved analyses stay as they are.
  • The version before can be put back from About.
  • From 4.25 or older: download MSpektra-5.12.zip below and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.12.zip below.

MSpektra 5.11

Choose a tag to compare

@its-kietsu its-kietsu released this 02 Nov 09:26

MSpektra 5.11

A new, much faster MSpektra. The windows, tools and files are the same as in 4.59.

New in 5.11

  • A peak under the description at the top of a spectrum keeps its label. The label moves just below the description.

New in 5.1

  • Peak labels: the Label tool and the right-click menu remove any label, also the automatic ones.
  • Display > Peak labels turns the automatic labels on or off.
  • A tool button clicked again turns the tool off.

New in 5.0

  • Much faster: it starts, opens files and draws plots in a fraction of the time.
  • Much smaller: the download is 33 MB (was 309 MB), the folder 82 MB (was about 870 MB).
  • Bigger plots: thin title strips and small gaps give the plots the space; no label is cut.
  • After a deconvolution the chromatogram, the spectrum and the result fit the window; the mass table is below.
  • A tidy folder: MSpektra.exe, README, LICENSES and the folder app.

Updating

  • From 4.26 or newer (also 5.0 and 5.1): click Update in the bar. Your settings and saved analyses stay as they are.
  • The version before can be put back from About.
  • From 4.25 or older: download MSpektra-5.11.zip below and unpack it.

Licences

  • Deconvolution uses UniDec. Please cite Marty et al., Anal. Chem. 2015, 87, 4370.
  • Thermo .raw files are read with RawFileReader reading tool. Copyright © 2016 by Thermo Fisher Scientific, Inc. All rights reserved. By using MSpektra you agree not to pass the RawFileReader files on to others.
  • The Agilent and Waters readers (msvendor.dll) are under the GNU LGPL 3; their source is msvendor-src-5.11.zip below.