- Ensure that the alignment laser is level with the table by checking its height at two separate, far apart locations on the bench.
- Ensure that the alignment laser is collimated. Below is a table of wavefront radii of curvature at different distances from the laser as measured with a Shack-Hartmann wavefront sensor.
| Distance, cm | RoC, m | | 50 cm | -30 +/- 1 | | 75 cm | -35 +/- 2 | | 100 cm | -40 +/- 3 |
- Set the galvanometer voltage to 0V