- No COM port
- Requires 2025 MCL drivers in the iSIM_Backup folder
- Drivers are at
\\sb-nas1.rcp.epfl.ch\LEB\Lab_business\Instruments\iSIM\Drivers\Mad City Labs - 2017 drivers do not work with Windows 11 and higher
- Driver updates require writing to MCL
- Drivers are at
- Micro-Manager device adapter: https://micro-manager.org/MCL_MicroDrive
- XY stage: https://micro-manager.org/MCL_MicroDrive
- Pre-Init Property:
EncodersPresent:No
- Pre-Init Property:
- Z Piezo: https://micro-manager.org/MCL_NanoDrive
NOTE: The XY List feature in Micro-Manager using the MCL_MicroDrive device adapter requires the following:
To correctly use the XY List feature, including the Calibrate button, the positioning type must be set to Absolute on the Device/Property Browser. However, Calibrate on the Device/Property Browser (see below) calibrates correctly using either absolute or relative positioning.