Skip to content

Python Script without GUI

ta1_k edited this page Aug 24, 2026 · 3 revisions

Using vs_controller without the GUI

vs_controller.py (at GUI/src/gui/controller/vs_controller.py) is self-contained: its only runtime dependency is pyserial.
It can be imported and driven from any Python script or Jupyter notebook

Make sure that you follow the PCB initial calibration steps before first time usage. The calibration guide is in this page

  • Highly recommend using GUI for this calibration.

Dependencies

Install pyserial in the environment the notebook uses:

pip install pyserial

Importing

# Script sits next to a copy of the file
from vs_controller import VoltageSourceController, DataLog

Jupyter usage pattern

In a notebook, keep one VoltageSourceController alive across cells rather than reconnecting per cell. A typical layout is:

  • Cell 1 — import + vs = VoltageSourceController() + configure_output
    • connect + start_reading. Run once at the top of the session.
  • Middle cells — call send_sweep / send_steady and read samples.
  • Last cell — send_off / stop_reading / disconnect.

connect and disconnect open and close the serial port; doing them repeatedly forces the ESP through its 2.5 s cable-watchdog and drops whatever run is in flight.

Lifecycle

connect  →  start_reading  →  send_…  →  stop_reading  →  disconnect
  • connect opens the serial port and starts the 2 s cable-watchdog heartbeat.
  • start_reading spawns the background reader that parses data,… / calib,… lines and populates the per-channel ring buffer.
  • stop_reading closes any open sweep log and joins the reader.
  • disconnect stops the heartbeat and closes the port.

If you skip start_reading, no telemetry is parsed and has_new_data stays False. If you skip connect (opening serial.Serial yourself), the firmware zeros every channel after 2.5 s because no heartbeat is running.

Public API

Static helpers

  • VoltageSourceController.list_ports() → list[tuple[device, description]] — USB-looking devices sorted first (by VID + keyword score).
  • VoltageSourceController.guess_port() → str — best pick, or "" if nothing looks USB-ish.

Connection state

  • is_connected: bool (property)
  • connected_port: str (property)
  • active_sweep_channels() → list[int]

Command senders

All are fire-and-forget writes terminated with \0. Each returns False silently if the port is not open.

For Channel ID and corresponding channels on the PCB, look at this page

  • send_off(ch)
  • send_steady(ch, voltage, duration, time_unit) Runs channel ch at the constant DAC voltage. time_unit accepts the firmware's timer tokens ("Sec", "Min", "Hour") or "Inf", which runs indefinitely and ignores duration. "Inf" is the common case — use it unless you specifically want the firmware to auto-turn-off after a bounded interval.
  • send_sweep(ch, range_v, step_size_mv) — sweeps ±range_v in step_size_mv millivolt steps. range_v is clamped to vs.sweep_range_max_v (default 1.5 V) with a console warning if exceeded; override the attribute on the instance if you really need a wider sweep.
  • send_calibration(ch, r_1k, r_gain, dac_vref) — persisted in NVS on the ESP.
  • send_odr(ch, sps) — ADC output data rate.
  • send_command(raw: str) → bool — raw escape hatch.

Live data reads

The reader thread parses each data,… line into a DataLog and appends it to a bounded ring buffer per channel (deque(maxlen=4096)).

  • has_new_data(ch) → bool — set whenever a new sample has arrived since the last consume.
  • get_latest_sample(ch) → DataLog | None — returns the newest sample and clears the new-data flag. O(1); intended for the main consumer loop.
  • peek_latest_sample(ch) → DataLog | None — same, but does not touch the flag. Safe from auxiliary threads.
  • get_channel_data(ch) → list[DataLog] — full copy of the ring. Pay this cost only when you need the whole history (e.g. at sweep end).
  • clear_channel_data(ch) — wipe the ring. Call before a fresh sweep so points from a prior run do not linger.

DataLog fields:

channel_id: int
mode: int        # 0=SWEEP, 1=STEADY, 2=IDLE, 3=CALIBRATION, 4=ODR
voltage: float   # V
current: float   # µA
time_s: float    # seconds since firmware boot (NOT wall time)

Example — Jupyter setup cell

Run this once at the top of the notebook. output/sweep/ holds sweep TXTs; output/log/ holds the diag file.

from vs_controller import VoltageSourceController

vs = VoltageSourceController()
vs.log_callback = lambda s: print("[esp]", s)

# Route sweep vs. diag/steady to separate subfolders. Call before connect
# so the diag log lands in output/log/ too.
vs.configure_output("output", sweep_subdir="sweep", log_subdir="log")

vs.connect(<Your USB connection>, 115200)
vs.start_reading()   

Teardown cell (run once when done):

vs.stop_reading()
vs.disconnect()

Example — steady

Prints each new sample — swap the print for a plot append, a variable capture, or nothing at all if you only care that the bias is on.

import time

ch, applied_v = 0, 0.25
vs.send_steady(ch, applied_v, 0, "Inf")

try:
    while vs.is_connected:
        s = vs.peek_latest_sample(ch)
        if s is not None:
            print(f"{s.voltage:.6f} V, {s.current:.6f} µA, {s.voltage / s.current} Ohm")
        time.sleep(1.0)   # sampling cadence
except KeyboardInterrupt:
    pass
finally:
    vs.send_off(ch)       

Example — sweep

Saves to user specified txt file name. Sweep range is software capped at $\pm 1.5V$

import time
import matplotlib.pyplot as plt

ch = 0
vs.clear_channel_data(ch)                              # discard prior samples
vs.start_sweep_log(ch, filename="sweep_A_1p5V.txt")    # optional TXT → output/sweep/
vs.send_sweep(ch, range_v=1.5, step_size_mv=10)

try:
    # Wait for the firmware to report OFF (mode == 2) after finishing the sweep.
    while vs.is_connected:
        s = vs.peek_latest_sample(ch)
        if s is not None and s.mode >= 2:
            break
        time.sleep(0.1)

    data = vs.get_channel_data(ch)          # list[DataLog], in arrival order
    voltages = [d.voltage for d in data]    # or: np.array([...])
    currents = [d.current for d in data]

    plt.plot(voltages, currents)
    plt.xlabel("Voltage (V)")
    plt.ylabel("Current (µA)")
    plt.show()
finally:
    vs.stop_sweep_log(ch)
    vs.send_off(ch)

Skip start_sweep_log / stop_sweep_log entirely if you don't want a TXT on disk — get_channel_data(ch) still works

vs stays connected between cells; only the teardown cell calls stop_reading / disconnect.

Output folder layout

  • configure_output(output_dir="output", sweep_subdir=None, log_subdir=None) — set where sweep TXTs and the diag log go. When both *_subdir args are None (the GUI default) everything lands flat in output_dir. Pass e.g. sweep_subdir="sweep", log_subdir="log" to split them into output/sweep/*.txt and output/log/*.log. Safe to call before connect — do it early if you want the diag log routed into log_subdir from the start.
  • start_reading accepts the same three kwargs and forwards them to configure_output, so a notebook can do it all in one call.

Sweep TXT logging

Automatic once start_reading is running: every data,… line with mode == 0 is written to the channel's open sweep file.

  • start_sweep_log(ch, filename=None) → path — auto-rotates by timestamp when filename is None. A bare filename is placed under the configured sweep dir; an absolute path is used as-is. Useful in notebooks where you want a run-specific name ("cooldown_A_1p2V.txt") rather than a timestamp.
  • stop_sweep_log(ch)
  • sweep_log_path(ch) → str

The start_steady_log / write_steady_row helpers are shaped around the GUI's meter-poll workflow (hard-coded pcb_current_uA + per-meter columns, row written only when the caller explicitly hands over a meter reading). For a pcb-only headless recording, ignore those helpers and stream peek_latest_sample(ch) yourself — see the steady example below.

Callbacks

Optional. Both default to None.

  • vs.log_callback = fn(msg: str) — every non-data, line from the firmware (boot banners, ESP_LOGx, printf, watchdog messages).
  • vs.calib_callback = fn(ch, r_1k, r_gain, dac_vref) — fires when the firmware sends a calib,… line (in reply to send_calibration, and on every boot).

Threads

All threads are daemon threads and live entirely inside vs_controller — no external ticker is required.

  1. Reader thread (start_reading) — parses data,… and calib,…; forwards everything else to log_callback; writes sweep rows.
  2. Heartbeat thread (connect) — writes hb every 2 s. The firmware zeros every channel after 4s without input. connect starts it, disconnect stops it.
  3. Reconnect thread — appears only after a USB drop, polls list_ports() for the last-known device, and re-runs connect + start_reading when it reappears. Bounded by _reconnect_timeout_s = 30 s. Set vs._auto_reconnect = False before disconnect() to opt out.

A diagnostic file vs_diag_<timestamp>.log is opened on each connect. It lands in the configured log dir (output_dir/log_subdir if you passed log_subdir, otherwise flat output_dir) and captures every non-data, line and heartbeat scheduling gaps. Line-buffered so it survives a killed process.

Somethings to be careful

  • connect alone does not spawn the reader — you must call start_reading.
  • send_command returns False silently if the port is not open. Check the return value or is_connected if you care.
  • DataLog.time_s is a firmware clock, not epoch. Stamp time.time() yourself when you read if you need wall time.
  • The class is thread-safe for the exposed operations. Do not reach into _serial or _data_buffers directly — use the accessors above.