Repository navigation
OpenOSC 0.20.2
Warning
Work in progress, for enthusiasts only. ZT703S is the main development and hardware test device.
Hardware coverage for the ZT706S is limited.
Install at your own risk. You can go back to the stock firmware.
What's new
Bode Analyzer coupling (#87)
The sweep sets both channels to AC coupling automatically. The scope restores the original coupling for each channel when the sweep completes, stops, or closes.
USBTMC with ngscopeclient (#91)
USBTMC now accepts large host read limits and returns the available reply. This fixes the stalled reads reported when adding a ZT703S in ngscopeclient over Linux USBTMC.
Scope V/div in Normal and Single (#93)
F1 Up/Down steps through the selected channel's standard range ladder. The held trace redraws at the selected scale while keeping its sample data. RUN uses the selected range for the new capture. When WAIT still shows an older trace, the F1 title shows the next range and the value row shows the plotted scale.
If an earlier version saved a digital gain setting that makes the scale look unexpected, set the affected channel's digital gain to 100% under Analyze → Display → View.
Consecutive screenshots (#94)
The confirmation reads Image saved as SCR00001.BMP using the saved file's name. The previous confirmation is cleared before the next screenshot is captured.
Boot credits
Added DivingDuck to the donors list. Thanks to @taylora1-1, @mb340, and @Nikonist for the reports and test details.
Testing
Installed 0.20.2 through Recovery on the ZT703S and verified the running version over SCPI. On the ZT703S, verified automatic AC coupling and restoration, large USBTMC reads and waveform transfer, F1 V/div changes in Normal and Single, and three consecutive BMP files. Both models' firmware and Recovery packages, native regression tests, and stack-budget checks passed.
Files
| Device | Recovery bootloader | Firmware |
|---|---|---|
| ZT703S | OpenOSC-Recovery-ZT703S.ZTK |
OpenOSC-Firmware-ZT703S.OSC |
| ZT706S | OpenOSC-Recovery-ZT706S.ZTK |
OpenOSC-Firmware-ZT706S.OSC |
Download the files for your model.
Updating
- From 0.3.0 or newer — Hold F2 at startup to open Recovery. Copy the matching
.OSCinto/openosc/UPDATE/. Select UPDATE FIRMWARE, choose the image, select CONFIRM, and press MENU. When the update completes, select START APPLICATION. - Recovery packages are byte-for-byte identical to those in 0.20.1. For this update, keep your installed Recovery if it is already at that version.
- From 0.2.0 or older, or for a new installation — Follow the installation instructions.
Known issues
- In capture configurations with a single slot, arming a capture can reuse the previous record's memory. WAIT then shows
No record to move. Completed records can be panned in HOLD. - Settings are written during an orderly power-off, restart, or entry to Recovery. A loss of power can discard changes made since the last save.
- USB data requires the 200/100 performance profile. USB hubs can cause disconnects.
- On the ZT703S, CH1 interleave applies DC offset correction by default. Run Interleave gain & timing to calibrate gain and timing for the unit.
SHA-256
b93d1bc713cca6a59184906dbe458a73130adb7994cfa9047077c44ef24b32e7 OpenOSC-Firmware-ZT703S.OSC
b5f5f09373a0ccb0ada4d7436691ce6d97da8358bc96e1e9b0435fad4e001b45 OpenOSC-Firmware-ZT706S.OSC
a389038d3aeec07168e0cc07d5f311815ca1054bf2f7c45a2c7b168680862d81 OpenOSC-Recovery-ZT703S.ZTK
6f1db91d198f5c406676c9e5751b1e73a69557c21dd63721c6a79180eb876fd2 OpenOSC-Recovery-ZT706S.ZTK