OpenOSC 0.20.3

Choose a tag to compare

@bve bve released this 22 Nov 22:18
· 10 commits to main since this release

Warning

Work in progress, for enthusiasts only. ZT703S is the main development and hardware test device.
Hardware coverage for the ZT706S is limited.
Install at your own risk. You can go back to the stock firmware.

What's new

Waveform density (#85)

Open Channel → Density for either channel in Y–T or Y–T+FFT. Colors marks less frequent paths in cooler colors and more frequent paths in warmer colors. Mono uses brightness in a selectable color. Density defaults to Off.

Use Analyze → Display → History → Persistence to set the history length. Density uses a one-second history when persistence is off. Infinite keeps observed paths, including rare events, until you clear history or change acquisition settings that reset it. HOLD freezes the display.

Scope digital gain steps (#98)

Each channel now steps from 1% through 25%, 50%, 75%, and 100% and continues in steps of 25 percentage points. An older off-grid value moves to the next step in the direction of the adjustment. The positive minimum keeps the volts-per-division scale defined.

Thanks to @Nikonist for the feature request and the digital gain report.

Testing

Installed 0.20.3 through Recovery on a ZT703S and confirmed the running version over SCPI. Also checked finite fading, Infinite retention of an occasional pulse through 272 ordinary acquisitions, the frozen HOLD display, and a return to exactly 100% gain on both channels. Both models' firmware and Recovery builds, native scope regression tests, and stack-budget checks passed.

Full changelog

Files

Device Recovery bootloader Firmware
ZT703S OpenOSC-Recovery-ZT703S.ZTK OpenOSC-Firmware-ZT703S.OSC
ZT706S OpenOSC-Recovery-ZT706S.ZTK OpenOSC-Firmware-ZT706S.OSC

Download the files for your model.

Updating

  • From 0.3.0 or newer — Hold F2 at startup to open Recovery. Copy the matching .OSC into /openosc/UPDATE/. Select UPDATE FIRMWARE, choose the image, select CONFIRM, and press MENU. When the update completes, select START APPLICATION.
  • Both Recovery packages match 0.20.2 byte for byte. If you already have Recovery from that release, only the firmware update is needed.
  • From 0.2.0 or older, or for a new installation — Follow the installation instructions.

Known issues

  • Dense two-channel captures with waveform density enabled can reduce the display frame rate.
  • In capture configurations with a single slot, arming a capture can reuse the previous record's memory. WAIT then shows No record to move. Completed records can be panned in HOLD.
  • Settings are written during an orderly power-off, restart, or entry to Recovery. A loss of power can discard changes made since the last save.
  • USB data requires the 200/100 performance profile. USB hubs can cause disconnects.
  • On the ZT703S, CH1 interleave applies DC offset correction by default. Run Interleave gain & timing to calibrate gain and timing for the unit.

SHA-256

aa7497302987083372fa10aba11ec6a28de8fd1e5539fec8948093f77787177a  OpenOSC-Firmware-ZT703S.OSC
886d5588c234d1138ca83dad4f64787b774e80a6e1b59916255debd27cd94944  OpenOSC-Firmware-ZT706S.OSC
a389038d3aeec07168e0cc07d5f311815ca1054bf2f7c45a2c7b168680862d81  OpenOSC-Recovery-ZT703S.ZTK
6f1db91d198f5c406676c9e5751b1e73a69557c21dd63721c6a79180eb876fd2  OpenOSC-Recovery-ZT706S.ZTK