OpenOSC 0.21.14

Choose a tag to compare

@bve bve released this 25 Nov 00:07
· 23 commits to main since this release

Warning

OpenOSC is in active development. ZT703S is the main hardware test device. Hardware coverage on ZT706S is limited. Install at your own risk. Return to the stock firmware.

What's new

Scope gain calibration

The gain calibration wizard now asks you to disconnect the signal and touch the selected probe tip to GND before measuring zero. If the input is outside the zero window, unstable or clipped, the wizard shows the reason and offers Retry ground. The retry starts a fresh measurement. Apply the reference when the wizard prompts for it.

Thanks to @Nikonist for reporting the sequence after channel matching. Issue #142

Testing

The native regression test saves a channel-match correction, then runs gain calibration through rejected ground readings with a DC signal, clipping and noise. After a valid ground reading, the test applies the simulated 3.3 V reference and verifies the saved gain. The firmware and Recovery builds, native update tests and stack checks passed for both board models.

Full changelog

Files

Device Recovery bootloader Firmware
ZT703S OpenOSC-Recovery-ZT703S.ZTK OpenOSC-Firmware-ZT703S.OSC
ZT706S OpenOSC-Recovery-ZT706S.ZTK OpenOSC-Firmware-ZT706S.OSC

Download the files for your model.

Updating

  • From 0.3.0 or newer — Hold F2 at startup to open Recovery. Copy the matching .OSC file to /openosc/UPDATE/. Select UPDATE FIRMWARE, choose the image, select CONFIRM, and press MENU. When the update finishes, select START APPLICATION.
  • The Recovery packages match 0.21.13 byte for byte. If Recovery is already installed, update the firmware alone.
  • From 0.2.0 or older, or for a new installation — Follow the installation instructions.

Known issues

  • USB data requires the 200/100 performance profile. USB hubs can cause disconnects.
  • Continuous waveform-envelope transfers scan the entire record and can reduce the update rate with deep memory. When requesting fresh frames at slow timebases, set the host's read timeout to more than the configured record duration plus one second.
  • CH1 interleaving is experimental on ZT703S. Running both ADCs at 280 MS/s can add noise. Offset, gain and timing calibration are available on-device.
  • Dense two-channel captures with waveform density enabled can reduce the display frame rate.
  • A single-slot capture can reuse the previous record's memory. If that happens while waiting for a trigger, panning becomes available again once a new record is captured.
  • Settings are saved on orderly power-off, restart or entry into Recovery. Unexpected power loss can discard changes made since the last save.

SHA-256

2e15e6aca2ff0f497837726fefb032a7aeb3a4bb1100071401c7f8ffbc438a67  OpenOSC-Firmware-ZT703S.OSC
fabf05d324d71a2650dcb1d5278285da8b120fa22994e902e517fdcd94758bfc  OpenOSC-Firmware-ZT706S.OSC
a389038d3aeec07168e0cc07d5f311815ca1054bf2f7c45a2c7b168680862d81  OpenOSC-Recovery-ZT703S.ZTK
6f1db91d198f5c406676c9e5751b1e73a69557c21dd63721c6a79180eb876fd2  OpenOSC-Recovery-ZT706S.ZTK