Skip to content

OpenOSC 0.21.17

Choose a tag to compare

@bve bve released this 26 Nov 03:28
· 10 commits to main since this release

Warning

OpenOSC is in active development. ZT703S is the main hardware test device. Hardware coverage on ZT706S is limited. Install at your own risk. Return to the stock firmware.

What's new

RS-232 and TTL serial decoding

Select RS-232 or UART (TTL) under Analyze → Decoder. RS-232 uses negative idle and an independent threshold that defaults to 0 V. Both modes support autobaud, configurable framing, and CSV export. Over SCPI, RS-232 is reported as RS232. Issue #83

Left arrow in menus

Settings → Display → Left key in menus offers Back (the default) and Step. In Step mode, Left selects the previous option or turns a switch off. In open lists, Left and Right move the selection without wrapping. F4 goes back or cancels. Thanks to @fletto2 for the original proposal. Issue #160 · PR #71

Multimeter reading alignment

The digital multimeter aligns numeric readings to the right and displays OL in the centre. Short readings and changes of sign keep the same right edge. Issue #162

Testing

The firmware and Recovery builds passed for both models. On a ZT703S, checked TTL and negative-idle serial decoding with CSV export, the menu shortcut including persistence across restart, and the transition from a number to OL and back. Native checks covered decoder protocols, the multimeter and menu widgets, settings and SCPI.

Full changelog

Files

Device Recovery bootloader Firmware
ZT703S OpenOSC-Recovery-ZT703S.ZTK OpenOSC-Firmware-ZT703S.OSC
ZT706S OpenOSC-Recovery-ZT706S.ZTK OpenOSC-Firmware-ZT706S.OSC

Download the files for your model. The Recovery images are identical to those in 0.21.16.

Updating

  • From 0.3.0 or newer — Hold F2 at startup to enter Recovery. Copy the matching .OSC to /openosc/UPDATE/. Select UPDATE FIRMWARE, choose the file, and confirm. After the update, select START APPLICATION.
  • From 0.2.0 or older, or for a new installation — Follow the installation instructions.

Known issues

  • The PNG encoder shares temporary acquisition memory with Auto Setup. Saving can be temporarily unavailable while that memory is in use.
  • USB data requires the 200/100 performance profile. USB hubs can cause disconnects.
  • Continuous waveform-envelope transfers scan the entire record and can reduce the update rate with deep memory. For fresh frames at slow timebases, set the host's timeout to more than the configured record duration plus one second.
  • CH1 interleaving is experimental on ZT703S. Running both ADCs at 280 MS/s can add noise. Offset, gain and timing calibration are available on-device.
  • Dense two-channel captures with waveform density enabled can reduce the display frame rate.
  • A single-slot capture can reuse the previous record's memory. If that happens while waiting for a trigger, panning becomes available again when a new record arrives.
  • Settings are saved on orderly power-off, restart or entry into Recovery. Unexpected power loss can discard changes made since the last save.

SHA-256

5b879d534d113d236ee5a51d8dff68c54956e556019e64c414ff83cbcf7a45c5  OpenOSC-Firmware-ZT703S.OSC
a3d6b5b7d1e73a7e7ac1bc78645958e72cf395a24bd5bb10cb89e90d63a44a7c  OpenOSC-Firmware-ZT706S.OSC
a9b9a8c120b0a045fc14ed698d75bc3b826242c7a816dfe6112c5606d8ff5113  OpenOSC-Recovery-ZT703S.ZTK
51ed387d80c88c9d36bf34b495b3cef10c90b1cea03c7eb55bed55dc3aa151cc  OpenOSC-Recovery-ZT706S.ZTK