Repository navigation
Warning
OpenOSC is in active development. ZT703S is the main hardware test device. Hardware coverage on ZT706S is limited. Install at your own risk. Return to the stock firmware.
What's new
Scope trigger types
The Scope Trigger menu now offers Pulse width, Runt, and Rise/fall time alongside Edge. Set the polarity or slope, voltage thresholds and time condition on the instrument or through SCPI. Normal and Single re-arm after a rejected candidate. Issue #133 · Trigger setup and SCPI
Multimeter session logging
Open MENU → Measurement → Session log to record interval statistics or log events when readings move outside a configurable band. The instrument saves CSV sessions and can reopen them as a trend with a cursor, zoom and pan. The file records elapsed time since the session started. Issue #144 · Session logging guide
Waveform CSV, decoder export and RS-232
Analyze → Save waveform CSV counts the exact file size when the initial estimate exceeds available space. If the file cannot fit, the instrument shows the required and free space. Decoder → Save CSV reports why a completed decode is unavailable, for example when a source channel is off. The decoder's F1 label now reads Protocol. RS-232 Auto hysteresis filters noise near the default 0 V threshold and applies to baud detection as well. Issue #164
Testing
Firmware and Recovery builds passed for both models. Host checks covered trigger capture, waveform CSV sizing and decoding. Testing on the ZT703S covered Pulse width, Runt, and Rise/fall time on both channels, interval session export and playback, waveform CSV and decoder CSV, and RS-232 with a near-zero input and a test pattern. Full changelog
Files
| Device | Recovery bootloader | Firmware |
|---|---|---|
| ZT703S | OpenOSC-Recovery-ZT703S.ZTK |
OpenOSC-Firmware-ZT703S.OSC |
| ZT706S | OpenOSC-Recovery-ZT706S.ZTK |
OpenOSC-Firmware-ZT706S.OSC |
Download the files for your model. The Recovery images are byte-for-byte identical to those in 0.21.18.
Updating
- From 0.3.0 or newer — Hold F2 at startup to enter Recovery. Copy the matching
.OSCto/openosc/UPDATE/. Select UPDATE FIRMWARE, choose the file and confirm. After the update, select START APPLICATION. - From 0.2.0 or older, or for a new installation — Follow the installation instructions.
Known issues
- The PNG encoder shares temporary acquisition memory with Auto Setup. Saving can be temporarily unavailable while that memory is in use.
- USB data requires the 200/100 performance profile. USB hubs can cause disconnects.
- Continuous waveform-envelope transfers scan the entire record and can reduce the update rate with deep memory. For fresh frames at slow timebases, set the host's timeout to more than the configured record duration plus one second.
- CH1 interleaving is experimental on ZT703S. Running both ADCs at 280 MS/s can add noise. Offset, gain and timing calibration are available on-device.
- Dense two-channel captures with waveform density enabled can reduce the display frame rate.
- A single-slot capture can reuse the previous record's memory. If that happens while waiting for a trigger, panning becomes available again when a new record arrives.
- Qualified triggers need the whole pulse or transition in the captured samples. The instrument has an interval between acquisitions when it cannot observe a new event.
- Counting the exact CSV size for a deep waveform can take several seconds. Use RAW when the full record exceeds the storage capacity for CSV.
- Settings are saved on orderly power-off, restart or entry into Recovery. Unexpected power loss can discard changes made since the last save.
SHA-256
d4cab101816a2c883d0f95ea78f6ba340b8e230ab4511521c5a72e80a382b336 OpenOSC-Firmware-ZT703S.OSC
b5f258f2f6ae7368124a8a52ee108cfc7df53e7f680595d93ab2324880f53054 OpenOSC-Firmware-ZT706S.OSC
a9b9a8c120b0a045fc14ed698d75bc3b826242c7a816dfe6112c5606d8ff5113 OpenOSC-Recovery-ZT703S.ZTK
51ed387d80c88c9d36bf34b495b3cef10c90b1cea03c7eb55bed55dc3aa151cc OpenOSC-Recovery-ZT706S.ZTK