OpenOSC 0.21.3
Warning
Work in progress, for enthusiasts only. ZT703S is the main development and hardware test device.
Hardware coverage for the ZT706S is limited.
Install at your own risk. You can go back to the stock firmware.
What's new
Scope grid and trace brightness (#86)
Use Analyze → Display → View → Grid brightness for the grid (0–100%) and Channel → Trace brightness for each channel (10–100%). All three default to 100% and adjust in 10% steps. Trace brightness also applies to Density, Glow, Persistence, Peak Hold and the XY trace (CH1). Channel labels and trigger markers keep their contrast.
Thanks to @OlegMedvedcuk for the suggestion.
Auto power off on battery (#104)
The external-power status pin can stay high after USB disconnects. The idle timer now runs in standalone mode despite that pin state. An active USB host or work such as generator output or recording still keeps the instrument on. Charging from a USB supply without a data host allows the timer to run.
Screenshot names after clearing (#105)
New screenshots include a checksum of their pixels in the name, for example SCR00001-644DF03C.BMP. This gives different screenshots different names when the same sequence number is reused after clearing. The image browser continues to support older SCRnnnnn.BMP files.
Thanks to @StrontiumD for both reports.
Testing
Installed the release package through Recovery on a ZT703S and checked the running version and waveform transfer over USB. On the same device, feature tests covered independent brightness in HOLD and after restart, auto power-off with the idle timer brought close to its deadline over SWD, and reading new screenshots through Recovery on Linux after deletion and reboot. Both models passed application and Recovery builds, package validation and stack checks. Native tests covered the FAT screenshot workflow, timer boundaries and settings.
Files
| Device | Recovery bootloader | Firmware |
|---|---|---|
| ZT703S | OpenOSC-Recovery-ZT703S.ZTK |
OpenOSC-Firmware-ZT703S.OSC |
| ZT706S | OpenOSC-Recovery-ZT706S.ZTK |
OpenOSC-Firmware-ZT706S.OSC |
Download the files for your model.
Updating
- From 0.3.0 or newer — Hold F2 at startup to open Recovery. Copy the matching
.OSCinto/openosc/UPDATE/. Select UPDATE FIRMWARE, choose the image, select CONFIRM, and press MENU. When the update completes, select START APPLICATION. - Both Recovery packages match 0.21.2 byte for byte. If you already have Recovery from that release, only the firmware update is needed.
- From 0.2.0 or older, or for a new installation — Follow the installation instructions.
Known issues
- CH1 interleaving is experimental on the ZT703S. Extra noise can still be visible when both ADCs are active at 280 MS/s. On-device offset and gain/timing calibration is available.
- Dense two-channel captures with waveform density enabled can reduce the display frame rate.
- In capture configurations with a single slot, arming a capture can reuse the previous record's memory. WAIT then shows
No record to move. Completed records can be panned in HOLD. - Settings are saved at an orderly power-off, restart or Recovery entry. Changes made since the last save can be lost if power is interrupted.
- USB data requires the 200/100 performance profile. USB hubs can cause disconnects.
SHA-256
64005501eb27012ef71fdabb5850d9f073ef6463ab58eb673331e406f37a6bf8 OpenOSC-Firmware-ZT703S.OSC
3f4af48cc1a3512a08ee61d71f5c539fd10cc376a907210c8a8a5944b0e75d64 OpenOSC-Firmware-ZT706S.OSC
a389038d3aeec07168e0cc07d5f311815ca1054bf2f7c45a2c7b168680862d81 OpenOSC-Recovery-ZT703S.ZTK
6f1db91d198f5c406676c9e5751b1e73a69557c21dd63721c6a79180eb876fd2 OpenOSC-Recovery-ZT706S.ZTK